==Introduction==
==Introduction==
Electrodeposition is a commonly used technique in which current is applied to layer one material, usually a metal, onto another.[[Image:200px-1937-Wheat-Penny-Front-Back.jpg|thumb|left|Wikipedia's image of an all-copper penny from 1937]] Modern pennies may be the most pervasive example of electrodeposition though older ones tucked away in piggy banks are not. Pennies used to be made of pure or nearly pure copper, but in 1982 when the price of copper increased and an all-copper penny had more than a penny’s worth of copper in it, a new kind of cent coin was minted. Post-1982 pennies are made with an inexpensive metal (zinc) that is covered with a thin copper-coat. [[Image:EarFlaresWithCondors giltcopper.jpg|thumb|right|Pair of earflares with condors from Loma Negra, Pirua Valley, 2nd-3rd century, The Metropolitan Museum of Art photo credit:web.site@metmuseum.org.]]The coating is applied using the positive terminal of a power supply to oxidize the material on the anode (a lump of copper in the case of pennies), allowing it to react with anions in solution. At the cathode, the material gets reduced (i.e. gains electrons), regains its zero valence state and is deposited on the cathode surface (i.e. zinc if we’re still talking pennies). The electrolytes solution that carries the anode material to the cathode is often made from very nasty chemicals like cyanides since these are great conductors and prevent corrosion of the anode. An active search for “friendlier” electrolytes continues, as does optimization of “electroless” deposition. Even without batteries, which Volta invented in the 1800s, incredible artifacts with decorative patterns of metal are found among some ancient ruins ([http://www.lablaa.org/blaavirtual/publicacionesbanrep/bolmuseo/1996/jldi41/jldi08a.htm e.g. in Loma Negra Peru]). Replacement plating was used as much as [http://www.jstor.org/view/00778958/ap050032/05a00080/0 2000 years ago to inlay the reactive element copper with less reactive silver and gold].
The Transmission Electron Microscope (TEM) achieves its remarkable resolution by “illuminating” samples using an electron beam in a vacuum rather than using a conventional light source in air. Since the electron beam passes through the sample that is being examined, the sample must be sufficiently thin and sufficiently sturdy to be hit by electrons in a vacuum. It’s important to remember that many biological materials are damaged or destroyed by the incoming electrons and that the TEM can image only the species that survive this harsh treatment. The denser parts of the sample will absorb or scatter some of the electron beam, and it’s the scattered electrons or those that pass through the sample that are focused using an electromagnetic lens. This “electron shadow” then strikes a fluorescent screen, giving rise to an image that varies in darkness according to the sample's density. For samples that are amenable to TEM, this form of examination can allow observation of angstrom-sized objects and of cellular details down to near atomic levels.[[Image:Macintosh HD-Users-nkuldell-Desktop-TEMgrid.jpg|thumb|left|Copper TEM grid with carbon mesh, photographed at 60X magnification]]


Successful electrodeposition requires that the cathode surface be extremely clean. Hydrophobic oils only molecules thick can inhibit deposition and cathode surfaces must be extensively cleaned with solvents, detergents, sonication (which uses sound waves to remove dirt) and/or acidic baths. Technical standards for cleaning, testing and assessing electrodeposited materials are described by [http://www.astm.org/cgi-bin/SoftCart.exe/index.shtml?L+mystore+hoqt3793+1196134766  the American Society for Testing and Materials]. Their guide for cleaning metals prior to electroplating has directions for precleaning, intermediate and final-cleaning as well as trouble shooting! [http://www.astm.org/cgi-bin/SoftCart.exe/DATABASE.CART/REDLINE_PAGES/B322.htm?L+mystore+hoqt3793+1196136492].  
Samples are applied to a wafer-thin "grid" before being loaded into the TEM and placed under vacuum. The grid can be made of many kinds of materials. All have lines of a conductive metal, in our case copper, that disperse the electron beam and thereby help keep the sample from being blown to bits by the energy in the beam. A carbon mesh is strung between the metal lines. Once a sample has been applied to the grid, it's only the portions that come to rest on the carbon mesh can be visualized, along with any imperfections in the carbon mesh itself.


[[Image:ITOphage tangle.png]] Today you will start to assemble your electrochromic device by electrodepositing IrO2 nanowires that formed on the specialize M13 phage. The IrO2 wires have a slight negative charge. Consequently they can be directed toward the patterned ITO slide if the slide is attached to the positive terminal of a power supply. The circuit is completed by attaching a platinum mesh to the ground terminal of a power supply. After this deposition, you will have a tangle of nanowires adhering to the areas of your slide that were not etched away last time. Next time you will finish building the electrochromic device, adding the Li+-layer of electrolytes and a second, unpatterned ITO slide to serve as the other conductive surface.
Today you'll wash the nanowires you synthesized last time and visualize a sample of them by TEM. The rest of the nanowires will be stored in a mortar until next time, when you'll grind them with some carbon and teflon for assembly into a battery.
<center>
[[Image:Map2TEM.png|thumb|500px| map from John to TEM facilities]]
</center>


==Protocols==
==Protocols==
Today 1/2 the lab will start with the deposition of their IrOx and phage onto their patterned ITO slide. The other half will have the opportunity to visualize a portion of their solution by TEM.  
Before you begin today's protocols, be sure to make some notes in your lab notebook about the appearance of the solution of nanowires. The next steps are devised to collect and wash the CTAB off the nanowires. You'll end up with some concentrated nanowires to observe by TEM and also to dry for assembly into a battery next time.
 
===Part 1: Collect nanowires===
===Part 1: Phage deposition onto patterned slide===
#Vortex the 50 ml falcon tubes and then sonicate the nanowires for 2 minutes. This step will break up large clumps of nanowires that may have formed.
[[Image:Deposition set up.JPG|thumb|right|200px|How to set up the ringstand for the deposition of the nanowires onto the ITO slide.]]
#Centrifuge your 50 ml falcon tubes in the clinical centrifuge at 3K for 10 minutes.
#Clean the ITO slide by placing it into a 50 ml falcon tube with 1% Liquinox.  Sonicate for 2 minutes.  Remove the slide from the Liquinox solution with gloves or with tweezers and briefly rinse in Millipore H2O.  Place the slide in a 50 mL Falcon tube with methanol.  Sonicate for 2 minutes. Remove the slide and then let it air dry on a paper towel on the bench. Do not wipe the ITO surface.
#Pour off the supernatant and resuspend the pellet in 1 ml of sterile water. You may need to vortex to resuspend the nanowires in the water.
#Plasma clean the slide for 5 minutes.  One of the teaching faculty will show you how to use the plasma cleaner. 
#Move 0.1 ml of your nanowires to an eppendorf tube and pool with team using same concentration of Ag+. Use this pool of nanowires for Part 2 of today's protocol.
#While the slide is drying, set up the ringstand.  Look at the figure to see how it should be set up.  Hold the platinum electrode and the slide with metal tweezers which you can then fix using clamps. 
#Add 9 ml of 100% EtOH to the remainder of the nanowires for Part 3 of today's protocol. Note that Part 2 and Part 3 of today's protocol can be done in parallel.
#Fill the beaker with 50 IrOx solution and add 1 ml of the phage stock that the teaching faculty will provide. 
#Lower the platinum mesh into the solution.  Try to get as much of the platinum into the solution WITHOUT letting the tweezers touch the solution.
#Lower the ITO slide into the solution with the ITO side facing the platinum.  Make sure that your pattern is fully immersed in the solution (or else you won’t get deposition on your entire pattern), but do not let the tweezers touch the solution. [[Image:ECD_set_up_close_up.JPG|thumb|left|200px|Your platinum mesh and ITO slide should be parallel to each other.]]
#The platinum electrode should be about ½ inch away from the ITO slide with the ITO side facing the platinum.  Try to have the slide and the platinum parallel to each other.
#Before you turn on the power supply, turn the voltage and the current all the way down (counter clockwise). 
#Attach the slide and the platinum mesh to the power supply using alligator clips.  The negative should be hooked up to the platinum and the positive should be hooked up to the ITO.  Also, the negative should be grounded on the power supply.  See the figure. [[Image:Power_supply_set_up.jpg|thumb|left|400px|Make sure your connections match those in the graphic.]]
#Turn on the power supply and slowly turn the voltage up to about 7.5 V.  Keep an eye on the current.  The current should not be above 0.01 amps.  If the current is higher than that, turn down the current knob.  If the current knob is all the way down, turn down the voltage.  Use the multimeter to confirm that the charge is flowing by measuring the voltage difference between the two tweezers.
#After 10 minutes, look at the solution.  You should see a dark cloud collecting near the ITO slide.  Turn off the voltage and slowly lift the ITO slide out of the solution. Use a 1 mL pipette to mix the solution.
#Carefully return the ITO slide to the solution.  Make sure that the tweezers are not touching the liquid.  Turn the voltage back on.  Check the voltage again using the multimeter.
#After another 10 minutes, you should see a dark cloud collecting near the ITO slide.  Turn off the voltage and slowly lift the ITO slide out of the solution.  Look at the pattern.  You should see a homogenous covering across the patterned part of the slide. If it is not fully covered, put the slide back in the solution for another 10 minutes.  If it is covered, carefully set the slide, pattern side up, on a paper towel to dry for at least 1 hour. 
#Look at the slide and if there is any additional liquid on it, you can carefully wick away the liquid using the tip of a Kimwipe.


===Part 2: TEM sample preparation===
===Part 2: TEM sample preparation===
Each group should prepare a TEM grid. This will allow for some duplicates to be visualized in case the grid is damaged or different kinds of EM are being performed.
#Vortex the nanowire pool for 1 minute at least and immediately remove 10 ul of the nanowire suspension to place on the <b> silver/black side (not the copper-looking side) </b> of the TEM grid that you have balanced in the specialized tweezers. <b> Treat the grid with care and use the tweezers only on the edge to minimize damaging the delicate mesh. </b>[[Image:Macintosh_HD-Users-nkuldell-Desktop-tweezersTEMgrid.jpg|thumb|right|TEM grid balanced in tweezers, wrong side up!]]<br>
# Allow the nanowires to settle onto the grid undisturbed for 20'.
# Remove the droplet from the grid by touching the edge with a Kimwipe thereby wicking the solution off the grid.
# Wash the grid by adding 10 uL of sterile H2O onto the grid. 
# Allow the grid to sit with water for 3 minutes and then wick dry.
# Wash the grid by adding 10 ul of 100% EtOH onto the grid.
# Allow the grid to sit with EtOH for 3 minutes and then wick dry, if it hasn't air dried.
# Place the grid into the TEM grid holder to transport to the [http://prism.mit.edu/finding/locatio.htm TEM facility (13-1012)]. A member of the teaching faculty will be in the facility to examine your samples with you.


The Transmission Electron Microscope (TEM) achieves its remarkable resolution by “illuminating” samples using an electron beam in a vacuum rather than using a conventional light source in air. Since the electron beam passes through the sample that is being examined, the sample must be sufficiently thin and sufficiently sturdy to be hit by electrons in a vacuum. It’s important to remember that many biological materials are damaged or destroyed by the incoming electrons and that the TEM can image only the species that survive this harsh treatment. The denser parts of the sample will absorb or scatter some of the electron beam, and it’s the scattered electrons or those that pass through the sample that are focused using an electromagnetic lens. This “electron shadow” then strikes a fluorescent screen, giving rise to an image that varies in darkness according to the sample's density. For samples that are amenable to TEM, this form of examination can allow observation of angstrom-sized objects and of cellular details down to near atomic levels.[[Image:Macintosh HD-Users-nkuldell-Desktop-TEMgrid.jpg|thumb|left|Copper TEM grid with carbon mesh, photographed at 60X magnification]]
===Part 3: Harvesting the remainder of your nanowires===
 
#Pool your sample with any group experimenting with the same ratio of gold and silver as you have tested.  
Samples are applied to a wafer-thin "grid" before being loaded into the TEM and placed under vacuum. The grid can be made of many kinds of materials. All have lines of a conductive metal, in our case copper, that disperse the electron beam and thereby help keep the sample from being blown to bits by the energy in the beam. A carbon mesh is strung between the metal lines. Once a sample has been applied to the grid, it's only the portions that come to rest on the carbon mesh can be visualized, along with any imperfections in the carbon mesh itself.  
#Centrifuge the sample of nanowires in EtOH using the clinical centrifuge (3K 10').  
# Mix the phage solution you prepared on the first day of this module with an equal volume of IrOx solution. These volumes will be on the order of 100 ul each.  
#Decant the supernatant and add 20 mL of sterile water and resuspend the centrifuged materials by vortexing if necessary.  
# Place 15 ul of your phage and IrOx solution on the <b> shiny, bright side </b> of the TEM grid that you have balanced in the specialized tweezers. <b> Treat the grid with care and use the tweezers only on the edge to minimize damaging the delicate mesh. </b>[[Image:Macintosh_HD-Users-nkuldell-Desktop-tweezersTEMgrid.jpg|thumb|right|TEM grid balanced in tweezers]]<br>
#Harvest using the clinical centrifuge (3K 10').
# Allow the nanowires to settle onto the grid undisturbed for 30'.
#Decant the supernatant and add 20 ml of 100% EtOH to resuspend the nanowires.  
# Remove the droplet from the grid with your P200 set to 50 ul.
#Harvest the nanowires in the clinical centrifuge one last time (3K 10').
# Wash the grid by adding 15 uL of sterile H2O onto the grid. Immediately remove the water.
#Decant the supernatant and wash the nanowires into a mortar with 100% EtOH.
# Dry the grid by very gently touching the edge of the grid to a piece of blotting paper in a petri dish.  Place the grid into the TEM grid holder to transport to the [http://prism.mit.edu/finding/locatio.htm TEM facility (13-1012)]. A member of the teaching faculty will be in the facility to examine your samples with you.
#Allow the nanowires to settle and remove the clear solution above them with a 1 ml pipetman.  
#Loosely cover the mortar with aluminum foil and set the covered mortar in the 30° incubator until next time. The majority of the EtOH will evaporate at 30°, though a vacuum oven will be used to fully dry the nanowires for next time.




Line 45: Line 47:


==For next time==
==For next time==
[[Image:ITOboundphage.png|thumb|left|image provided by Yoon Sung Nam]]
You should now be considering the presentation materials themselves and if you can, start working on the materials you will use to describe your idea. Reconsult the [[20.109(F09): Mod 3 Day 6 Oral presentations| specific directions for what you'll need]] as well as the more general [[20.109(F09):Guidelines for oral presentations| guidelines for all oral presentations]].
 
# The deposition of the nanowires onto the patterned slide is driven by current, resulting in an uneven tangle of IrO<sub>2</sub>wires on the ITO. As a biological engineer, use your understanding of M13 biology to engineer a more densely packed and uniform surface of nanowires on the ITO, like the idealized cartoon shown in the image below.
# Your oral presentations of your research proposals/exhibit will be given one week from today. You should be well on your way to preparing the presentation materials themselves and should be working on the wording you will use to describe your idea. Reconsult the [[20.109(F08): Module 3 oral presentations| specific directions for what you'll need]] as well as the more general [[20.109(F08):Guidelines for oral presentations| guidelines for all oral presentations]].

Latest revision as of 05:18, 27 November 2009


20.109(F09): Laboratory Fundamentals of Biological Engineering

Home        People        Schedule Fall 2009        Assignments        Lab Basics        OWW Basics       
DNA Engineering        System Engineering        Biomaterials Engineering              

TEM

Introduction

The Transmission Electron Microscope (TEM) achieves its remarkable resolution by “illuminating” samples using an electron beam in a vacuum rather than using a conventional light source in air. Since the electron beam passes through the sample that is being examined, the sample must be sufficiently thin and sufficiently sturdy to be hit by electrons in a vacuum. It’s important to remember that many biological materials are damaged or destroyed by the incoming electrons and that the TEM can image only the species that survive this harsh treatment. The denser parts of the sample will absorb or scatter some of the electron beam, and it’s the scattered electrons or those that pass through the sample that are focused using an electromagnetic lens. This “electron shadow” then strikes a fluorescent screen, giving rise to an image that varies in darkness according to the sample's density. For samples that are amenable to TEM, this form of examination can allow observation of angstrom-sized objects and of cellular details down to near atomic levels.
Copper TEM grid with carbon mesh, photographed at 60X magnification

Samples are applied to a wafer-thin "grid" before being loaded into the TEM and placed under vacuum. The grid can be made of many kinds of materials. All have lines of a conductive metal, in our case copper, that disperse the electron beam and thereby help keep the sample from being blown to bits by the energy in the beam. A carbon mesh is strung between the metal lines. Once a sample has been applied to the grid, it's only the portions that come to rest on the carbon mesh can be visualized, along with any imperfections in the carbon mesh itself.

Today you'll wash the nanowires you synthesized last time and visualize a sample of them by TEM. The rest of the nanowires will be stored in a mortar until next time, when you'll grind them with some carbon and teflon for assembly into a battery.

map from John to TEM facilities

Protocols

Before you begin today's protocols, be sure to make some notes in your lab notebook about the appearance of the solution of nanowires. The next steps are devised to collect and wash the CTAB off the nanowires. You'll end up with some concentrated nanowires to observe by TEM and also to dry for assembly into a battery next time.

Part 1: Collect nanowires

  1. Vortex the 50 ml falcon tubes and then sonicate the nanowires for 2 minutes. This step will break up large clumps of nanowires that may have formed.
  2. Centrifuge your 50 ml falcon tubes in the clinical centrifuge at 3K for 10 minutes.
  3. Pour off the supernatant and resuspend the pellet in 1 ml of sterile water. You may need to vortex to resuspend the nanowires in the water.
  4. Move 0.1 ml of your nanowires to an eppendorf tube and pool with team using same concentration of Ag+. Use this pool of nanowires for Part 2 of today's protocol.
  5. Add 9 ml of 100% EtOH to the remainder of the nanowires for Part 3 of today's protocol. Note that Part 2 and Part 3 of today's protocol can be done in parallel.

Part 2: TEM sample preparation

Each group should prepare a TEM grid. This will allow for some duplicates to be visualized in case the grid is damaged or different kinds of EM are being performed.

  1. Vortex the nanowire pool for 1 minute at least and immediately remove 10 ul of the nanowire suspension to place on the silver/black side (not the copper-looking side) of the TEM grid that you have balanced in the specialized tweezers. Treat the grid with care and use the tweezers only on the edge to minimize damaging the delicate mesh.
    TEM grid balanced in tweezers, wrong side up!

  2. Allow the nanowires to settle onto the grid undisturbed for 20'.
  3. Remove the droplet from the grid by touching the edge with a Kimwipe thereby wicking the solution off the grid.
  4. Wash the grid by adding 10 uL of sterile H2O onto the grid.
  5. Allow the grid to sit with water for 3 minutes and then wick dry.
  6. Wash the grid by adding 10 ul of 100% EtOH onto the grid.
  7. Allow the grid to sit with EtOH for 3 minutes and then wick dry, if it hasn't air dried.
  8. Place the grid into the TEM grid holder to transport to the TEM facility (13-1012). A member of the teaching faculty will be in the facility to examine your samples with you.

Part 3: Harvesting the remainder of your nanowires

  1. Pool your sample with any group experimenting with the same ratio of gold and silver as you have tested.
  2. Centrifuge the sample of nanowires in EtOH using the clinical centrifuge (3K 10').
  3. Decant the supernatant and add 20 mL of sterile water and resuspend the centrifuged materials by vortexing if necessary.
  4. Harvest using the clinical centrifuge (3K 10').
  5. Decant the supernatant and add 20 ml of 100% EtOH to resuspend the nanowires.
  6. Harvest the nanowires in the clinical centrifuge one last time (3K 10').
  7. Decant the supernatant and wash the nanowires into a mortar with 100% EtOH.
  8. Allow the nanowires to settle and remove the clear solution above them with a 1 ml pipetman.
  9. Loosely cover the mortar with aluminum foil and set the covered mortar in the 30° incubator until next time. The majority of the EtOH will evaporate at 30°, though a vacuum oven will be used to fully dry the nanowires for next time.


DONE!