AU Biomaterials Design Lab:Equipment/LS55 PerkinElmer: Difference between revisions
Dhea Patel (talk | contribs) |
Dhea Patel (talk | contribs) |
||
Line 70: | Line 70: | ||
'''Scanning speed''' | '''Scanning speed''' | ||
*Increments: 1nm | *Increments: 1nm | ||
*Rate: 10-1500 nm/min | *Rate: 10-1500 nm/min<br> | ||
'''Sensitivity''' | '''Sensitivity''' | ||
*Signal to Noise: 500:1 rms | *Signal to Noise: 500:1 rms | ||
**using the Raman band of water with excitation at 350 nm | |||
** excitation and emission bandpass: 10nm<br> | |||
'''Standard Cellholder''' | |||
*A single position water thermostattable holder for 10mm cuvettes. | |||
'''Physical'''<br> | '''Physical'''<br> | ||
*Dimensions (W×D×H): 450(W) x 490(D) x 270(H) and Weight is 15kg<br> | *Dimensions (W×D×H): 450(W) x 490(D) x 270(H) and Weight is 15kg<br> |
Revision as of 07:54, 6 December 2012
Description
- Perkin Elmer LS 55 Fluorescence Spectrometer
- This is a 0.1nm resolution, compact spectrophotometer that uses the windows-compatible FL WinLab™ software that can measure fluorescence, phosphorescence and bio- and chemi-luminescence using a pulsed Xenon lamp source.
Purpose of use
Instrumentation The LS55 Perkin Elmer Fluorescence Spectrometer, 230V uses a pusled Xenon source to measure fluorescence, phosphorescence, chemi- and bio- luminescence. The FL WinLab™ software has different modes operation like Scan, TimeDrive, and Ratio Data Collection, increasing the versatility of the instrument. The excitation and emission monochromators can be used independently or simultaneously. The instrument also has the ability to prescan the sample to determine the excitation peak and maximum emission.
Included with the LS55 Perkin Elmer Fluorescence Spectrometer, 230V is a single-cell sample holder that can hold up to a 1cm pathlength cell and has the option of stirring. The automated polarizer has two filter wheels that have both a horizontal and vertical polarizing element. The polarizer is controlled by the software.
Applications
- Fluorescence, phosphorescence and bio- and chemi-luminescence
- 3D excitation/emission scans
- 3D synchronous and kinetic scans
- Flat samples, like TLC plates and electrophoresis gels, can be analyzed using the plate reader accessory
- Wavelength kinetics (both single and multiple wavelengths)
- Intracellular ion analysis
Location
- Beeghly 207
- American University
- Washington, DC
- access to the laboratory is restricted; see Dr. Miller, Dr. Hartings or Dr. Fox for access.
Contact
- contact with questions or to schedule access via email
Use
- How does one use this instrument.
- Include any relevant links to OWW protocols
- Include any relevant links to the manufacturer's protocols
- Note to users: you may note any troubleshooting or log your use on the talk page.
Safety
- Are there any safety considerations for this instrument?
Specifications
Spectral
- Wavelength range: 190 to 1100nm
- Spectral bandwidth: 1nm (190 to 1100nm)
- Wavelength accuracy: ±0.1nm
- Wavelength repeatability: ±0.5nm
Source
- Source: Xenon discharge lamp, equivalent to 20kW for 8μs duration
- Pulse Width at Half Height: <10μs
Detectors
- Sample Detector:
- Gated photomultiplier with modified S5 response, max operation ~650nm
- Red-sensitive R928 photomultiplier (optional), max operation ~900nm
- Reference detector: Photodiode, max operation ~900nm
Monochromators
- Monk-Gillieson type monochromators
- Excitation:
- Range: 200-800nm with zero order selectable
- Spectral bandpass: 2.5-15.0nm, selected in 0.1nm increments
- Emission:
- Range: 200nm-650nm with standard photomultiplier with zero order selectable; 200-900m with optional R928 photomultiplier
- Spectral bandpass:2.5-20.0nm, selected in 0.1nm increments
- Filters:
- Computer selectable cut-off (high pass) at 290, 350, 390, 430, and 515nm
- A blank that acts as a shutter
- A 1% T attenuator
- Clear beam position
- Excitation:
Phosphorescence Mode
- Variable delay and gate times
- Minimum total period: 12.0ms (50Hz)
Scanning speed
- Increments: 1nm
- Rate: 10-1500 nm/min
Sensitivity
- Signal to Noise: 500:1 rms
- using the Raman band of water with excitation at 350 nm
- excitation and emission bandpass: 10nm
Standard Cellholder
- A single position water thermostattable holder for 10mm cuvettes.
Physical
- Dimensions (W×D×H): 450(W) x 490(D) x 270(H) and Weight is 15kg
- Memory: USB memory, Saved as text and UVPC file
References
Useful links
Add any other categories that may be relevant to your page. We use categories like tags.