==Specifications== | ==Specifications== | ||
* Optics: Double beam, chopper mirror | |||
* Monochromator: Czerny-Turner monochromator, holographic grating (1600 lines/mm) | |||
* Wavelength range: 190-900 nm, automated wavelength selection | |||
* Slit width: 0.2 nm or 0.7 nm, manual selection | |||
* Lamp turret: 2 lamps, simultaneously lit, manual selection | |||
* Lamp mode: Emission, non-BGC, BGC-D<sub>2</sub> | |||
* Frequency: 100 Hz | |||
* Nebulizer: Nebulizer integrated impact bead and jacket tip, Pt/Ir capillary | |||
* Chamber: Polypropylene | |||
* Burner: Fixed position and height, 10 cm titanium slot burner, manual selection between air/C<sub>2</sub>H<sub>2</sub> and N<sub>2</sub>O/C<sub>2</sub>H<sub>2</sub> | |||
* Gas control: Manual selection of flow rate, automatic air/N<sub>2</sub>O switching system | |||
* Safety: Gas pressure monitoring, automatic flame monitoring, safety interlock for misuse of burner, automatic flame extinguish upon power failure | |||
* Ignition: Manual selection | |||
* Software: Based on Microsoft Windows, WizAArd | |||
* Power requirements: AC220 V, 230 V, 50/60 Hz, 300 VA, requires transformer for AC115 V | |||
* Ambient temperature range: 10~35°C | |||
* Ambient humidity range: 45~80%, but less than 70% if temperature is above 30°C | |||
==References== | ==References== |
Revision as of 13:24, 13 November 2012
Description
- Shimadzu AA-6200 double-beam atomic absorption spectrophotometer.
- It can detect the concentration of Au and Cu even at very low concentrations (ppm).
Purpose of Use
Location
Use
Safety
Specifications
- Optics: Double beam, chopper mirror
- Monochromator: Czerny-Turner monochromator, holographic grating (1600 lines/mm)
- Wavelength range: 190-900 nm, automated wavelength selection
- Slit width: 0.2 nm or 0.7 nm, manual selection
- Lamp turret: 2 lamps, simultaneously lit, manual selection
- Lamp mode: Emission, non-BGC, BGC-D2
- Frequency: 100 Hz
- Nebulizer: Nebulizer integrated impact bead and jacket tip, Pt/Ir capillary
- Chamber: Polypropylene
- Burner: Fixed position and height, 10 cm titanium slot burner, manual selection between air/C2H2 and N2O/C2H2
- Gas control: Manual selection of flow rate, automatic air/N2O switching system
- Safety: Gas pressure monitoring, automatic flame monitoring, safety interlock for misuse of burner, automatic flame extinguish upon power failure
- Ignition: Manual selection
- Software: Based on Microsoft Windows, WizAArd
- Power requirements: AC220 V, 230 V, 50/60 Hz, 300 VA, requires transformer for AC115 V
- Ambient temperature range: 10~35°C
- Ambient humidity range: 45~80%, but less than 70% if temperature is above 30°C