Avoiding RNase contamination

From OpenWetWare
Revision as of 16:26, 5 December 2006 by Reshma P. Shetty (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigationJump to search

Some guidelines on how to best avoid RNase contamination.

Common sources of contamination

  1. Buffers contaminated with microorganisms. (Note that the solutions need not be turbid to be problematic.)
  2. Pipettors. Keep a separate set of pipettors for RNA use to avoid contamination with RNases. Avoid touching the barrel or metal ejector to the sides of tubes.

Precautions

  1. Keep separate a set of pipettors, glassware, plasticware, reagents, electrophoresis apparatus etc. for RNA use only.
  2. Store solutions in small aliquots and discard each aliquot after use.
  3. Clean electrophoresis apparatus.
    1. Wash with detergent solution.
    2. Rinse in H2O.
    3. Dry with ethanol.
    4. Fill with 3% solution of H2O2.
    5. Incubate 10 mins at room temperature.
    6. Rinse with DEPC treated H2O.
  4. Handle chemicals handled with disposable spatulas.
  5. Whenever possible treat solutions with 0.1% DEPC for 1 hour at 37°C and autoclave 15 mins at 15 psi.
  6. Bake glassware at 300°C for 4 hours.
  7. Treat plasticware with RNAseZap or DEPC.
  8. Use RNase-free disposable tips and tubes. Use sterile forceps to transfer items to racks.

Reference

  1. Molecular Cloning