LAB 9 Analyzing Your DNA Sequencing data and Finishing the Project

You will have a workshop to evaluate and analyze your 16S rRNA gene sequencing data

Finishing the project

After you have recorded all your isolate test results on a lab Google Doc, please discard all cultures. You will receive 5 "clean-up points" if we do not find any evidence of your cultures or tests you will receive 5 points toward your lab grade. If we have to discard and clean-up later after you, you will not receive any of those points. The date that clean up must be completed will be announced by your instructor.

Identification of bacteria using MALDI-TOF Mass Spectronomy

Using Matrix Assisted Laser Desorption/Ionization Time of Flight Mass Spectrometry (MALDI-TOF MS) to Identify bacteria

Using log phase (24 hour) cultures of your isolates you will learn how the MALDI-TOF Biotyper works to identify your microbes using a different reference standard than our other ID method (comparison of 16s RNA gene sequences). This method compares particular proteins by mass spectrometry to reference standards. You need freshly grown cultures of your isolates today made 24 hours prior to this lab. If you have one or two very slow growing cultures, you need to allow enough time to have a small but visible colony on your plate.

BACKGROUND:
The following MALDI background information is provided courtesy of Bruker Daltronics:





Glossary
MALDI is the abbreviation for Matrix Assisted Laser Desorption/Ionization
TOF for Time of Flight.
Matrix is a chemical substance which will assist the sample ionization process and must be crystallized in a sample compound on the target surface. Common matrix substances are:

    • CCA or HCCA (Alpha-Cyano-4-Hydroxycinnamic Acid)
    • SA (Sinapinic Acid)
    • DHB (Dihydroxybenzoic Acid)

MALDI-TOF The samples which are measured for MALDI Biotyper application are typically the Proteins of Bacteria, Fungi and Yeast. The mass and distribution information will be used for identification.

By firing with a laser (on microflex a nitrogen uv laser with a wavelength of 337 nm) on the prepared matrix-sample compound the matrix will be vaporized. The sample will be entrained and ionized. The matrix and sample ions will be accelerated in an electrical field as the target plate is under high voltage of up to 20 kV against to an arranged ground potential.

Via the flight time between acceleration and hitting the detector the mass of an ion can be determined (TOF = Time of Flight technique). The complete process needs to take place under vacuum of better than 3*10-6 mbar. As better the vacuum as less collisions


To avoid influences of temperature, mechanical and electrical deviations and special techniques for acceleration the instrument needs to be calibrated before measuring unknown samples. To do this a mixture of known samples will be measured and an assignment between flight time and theoretical mass will be done.


The achievable resolution depends on the speed of the detector and the speed of converting the analog signals to digital signals. On microflex a 500 MHz digitizer which achieves a distance from 2 ns between data points (in 2 ns the light covers a distance of just 60 cm) is used. With a special digitizer function provides actually a calculated ‘pseudo’ 2 GHz sampling rate. Mainly influencing variables are also the applied laser energy, the drift distance and the electrical field distribution.

Definition and calculation of mass resolution
The mass resolution for a single peak is calculated by dividing the peak mass by the Full Width at Half Maximum (FWHM). The resolution value can be checked in flexControl and flexAnalysis directly.


Linear mode The ionized sample will be accelerated in an electrical field between the target plate and the acceleration electrodes. Followed by ‘drifting’ to a field free region and finished by hitting the detector.


Ion acceleration

Continuous extraction:
The first MALDI-TOFs have been used continuous extraction to accelerate ions. Between target plate and ground plate is a static electrical field which accelerates the ions immediately after generating.

Delayed extraction:
Next step was to use a delayed extraction to accelerate ions (pulsed ion extraction - PIE). Between target plate and ground plate a second potential plate (P2) is placed to modify the electrical field. The P2 plate has the same potential as the target plate. So no acceleration will happen after generating the ions. After a delay time, the P2 potential will be decreased and the acceleration starts. The fortune is that the effect of loosing resolving power because of different kinetic energy of equal ions will be reduced. A higher resolution for a small mass range can be observed.



A further advancement could be reached by optimizing the pulse shape applied to the P2 plate. With the now called PAN (panoramic) mode a high resolution over a brought mass range will be reached.


Electrical Lens:
The ion lens is located on ion optic behind the acceleration electrodes and applies a small vectored electrical field to focus the ion cloud for better resolution.

Connections:
The P1 (IS-1) potential (Target plate) is connected to the PIE electronic J497.
The P2 (IS-2) potential is connected to the PIE electronic J495.
The Lens potential is connected to the Lens power supply J480.

Procedure
The Bruker MALDI Biotyper requires organic acid solutions to: 1) prepare the Bacterial Standard (BTS) and HCCA (matrix): 2) complete the Formic Acid Extraction process; and 3) to clean the Polished Steel Target.
Preparing your isolates for MALDI-TOF MS
Formic Acid Isolate Extraction
Principle: In some microorganisms it may be necessary to break down the cell wall and separate the ribosomal proteins prior to spectra analysis.

The Tube Formic Acid Extraction procedure (TE)

  • Materials Needed :
    • Eppendorf 1.5 μL microfuge tubes
    • Eppendorf pipettes/tips
    • Microfuge
    • Toothpicks (or applicable transfer device)
    • Vortex
    • Microfuge tube rack
    • Bacterial Test Standard (BTS)
    • Target
    • Ultra-Pure Water, HPLC/MS Grade
    • Ethanol, 100% HPLC/MS Grade
    • Formic Acid, 70% HPLC/MS Grade
    • Acetonitrile, 100 % HPLC/MS Grade
    • Fresh Microorganism: overnight growth should be used for routine microorganism identification; slow-growing bacteria may need to cultivate for several days before testing; do not use organisms that have been stored at 40C or lower as this has a negative impact on quality of spectra and reproducibility; storing plates at room temperature for several days is acceptable; different media types and growth temperatures have little effect on results.

1. Add 300 μL of HPLC water to each Eppendorf microfuge tube
2. Transfer a large, single colony of microorganism to the tube (more than one colony may need to be transferred if microorganism is small you want the solution to be opaque/cloudy; chose isolated colonies OR, if using broths, spin and wash (HPLC water) 2-3x to remove media)); Vortex thoroughly (1) minute.
3. Add 900 μL of Ethanol; Vortex thoroughly (1) minute.
4. Tabletop Centrifuge at maximum (13,000 to 15,000 rpm) speed for (2) minutes
5. Decant Ethanol; Tabletop Centrifuge for (2) minutes high
6. Remove ALL excess Ethanol with pipette (tubes may be left at room temperature in the biological hood, or spun more than once, to complete the evaporation process if necessary, all the ethanol must be gone before the next step).
7. Add 50 μL of 70% Formic Acid (if only a small amount (size of pellet) of microorganism is available decrease the Formic Acid volume to 10 μL).
8. Vortex thoroughly 1 min and let stand for 5 minutes.
9. In the chemical hood and WEARING GLOVES, add 50 μL of 100% Acetonitrile (if only a small amount (size of pellet) of microorganism is available decrease the Acetonitrile volume to 10 μL); vortex thoroughly (1) min; NOTE: the volumes of 70% Formic Acid and Acetonitrile must be equal volumes. (See Formic Acid Extraction Solution 2.)
10.Centrifuge at maximum speed (13,000 to 15,000 rpm) for (2) minute –(continue with: MALDI Target Preparation).


Note: A Bacterial Test Standard (BTS) 10 mg/ml (w/v) is used to calibrate and test software, use fresh on every plate.
The BTS will be provided in lab ready to use.
alpha-Cyano-4-hydroxycinnamic acid (HCCA (MATRIX)) (from Bruker, Fluka, Sigma) Bruker Product number 255344:.
The reconstituted matrix will be provided in lab ready to use.

MALDI Target Preparation:
Materials Needed:

  • MALDI Biotyper Target
    • Bacterial Test Standard (BTS)
    • α-Cyano-4-hydroxycinnamic acid HCCA (matrix)
    • 2 μL Pipets
    • Eppendorf Brand Pipette/Tips
    • MALDI Biotyper Target Worksheet
    • Fresh Microorganism: overnight growth should be used for routine microorganism identification; slow-growing bacteria may need to cultivate for several days before testing. Do not use organisms that have been stored at 40C or lower, as this has a negative impact on quality of spectra and reproducibility; storing plates at room temperature for several days is acceptable. Different media types and growth temperatures have little effect on results. PREPARE CULTURES FOR MALDI USING FORMIC ACID EXTRACTION IN LAB.

TARGET LOADING:
1. Prepare a MALDI Target Worksheet with appropriate sample identification, assigning the first two target spots for BTS (QC) and the rest for each isolate to be analyzed.
2. Inoculate 1 µL of each Isolate Formic Acid extract supernatant (avoid touching the pellet) onto the target spots indicated on the Target Worksheet.
3. Once all the isolate extracts are loaded.
4. One student will place 1μL of BTS on each of the 2 BTS spots.
5. Air Dry all spots at Room temperature.
6. Immediately after BTS spots (and all other spots) are dry, overlay all spots with 1 µL of HCCA (Matrix) (to avoid oxidation of the standard). Once the HCCA is uncapped work quickly to avoid evaporation!!!
7. Be sure the target plate is completely dry before placing the Target plate in the MALDI-TOF.
8. You are ready to analyze your isolates using the MALDI-TOF MS.

Assignment

Study for your Lab Practical (45points)
Tips: Lab Practical.
Your final paper is due on Tues. Dec. 10 by 4pm. (the last day of classes) for all students. You have all the data you need to start working on it. Instructions can be found at: Assignment: Final Paper To Do:
Discard and clean up any remaining test materials and old cultures of your isolates. Note that there are 5 bonus points awarded for perfect clean-up. Your instructor will explain the criteria for obtaining these points.