BME100 f2013:W900 Group10 L3: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
Line 61: Line 61:
'''T-Test:'''
'''T-Test:'''


Analysis of the data with a t-test shows that there is a significant difference between data sets. Thus the p-value is well below the value required to achieve significance difference (0.05). However, we do not want a significant difference between the data from the oral thermometer and the data from the sensor. This indicates that the sensor is not accurate.
Analysis of the data with a t-test shows that there is a significant difference between data sets: The p-value is well below the value required to achieve significance (0.05). However, we do not want a significant difference between the data from the oral thermometer and the data from the sensor. This indicates that the sensor is not accurate.


'''Pearson's r:'''
'''Pearson's r:'''

Revision as of 11:37, 8 October 2013

BME 100 Fall 2013 Home
People
Lab Write-Up 1 | Lab Write-Up 2 | Lab Write-Up 3
Lab Write-Up 4 | Lab Write-Up 5 | Lab Write-Up 6
Course Logistics For Instructors
Photos
Wiki Editing Help


OUR TEAM

Name: Barrett Anderies
Name: Liam Williams
Role(s)
Name: Duran Charles
Role(s)
Name: Jolin Jose
Role(s)
Name: student
Role(s)
Name: student
Role(s)

LAB 3A WRITE-UP

Descriptive Statistics

In this experiment, two medical devices are used to collect body temperature data. The first device used was the RAIING Medical Company Wireless Thermometer (Model: WTM-BT30-T). The other medical device used was a standard oral thermometer, the accepted standard for reading body temperature, against which we will compare the RAIING Wireless Thermometer. The table below shows the average, standard deviation, end point #'s and standard errors of body temperature for each medical device. Note that some measurements were taken in the lab under controlled conditions (not moving, etc.). These data are labeled "Inside". The rest of the data was taken outdoors during activity, and these data are labeled "Outside".



Results





Analysis



Summary/Discussion

T-Test:

Analysis of the data with a t-test shows that there is a significant difference between data sets: The p-value is well below the value required to achieve significance (0.05). However, we do not want a significant difference between the data from the oral thermometer and the data from the sensor. This indicates that the sensor is not accurate.

Pearson's r:

The Pearson's r coefficient shows that the data has little linear correlation. This implies that that, when compared with the readings from the accepted measurement device (oral thermometer), the sensor reports widely varying and inconsistent temperature readings. Thus the sensor is imprecise.

Trend Line:

If the sensor were to report precise data, we would expect the slopes of the trend lines to be close to one. The slope of our trend lines are well below one, indicating that the sensor does not report precise data.

Standard Deviation:

If we make the assumption that our standard device (oral thermometer) is precise and that our sensor is too, then we would expect both data sets to have similar standard deviations. This is not the case, with the standard deviation being greater with the sensor data for both inside and outside data sets. This implies that our sensor reports data with more variation than the oral thermometer, indicating that the sensor is not precise.

Overall lack of accuracy

The sensor consistently reported body temperature lower than the oral thermometer (see scatter plots and t-test), indicating that, if we take our oral thermometer (our standard) to be accurate, the sensor is not accurate.

General Issues with the Sensor

While using the RAIING thermometer we encountered several design/function issues. First, we were unable to achieve wireless communication between the device and an iPhone, rendering the device useless for automatic data recording (we had to manually read off data from the on-device display). Second, the device was rather large, making it difficult to position on petite people. And third, the device had to be secured tightly to ensure proper contact with the skin, resulting in some restriction of bloodflow and skin irritation.

Summary

All means of analysis yielded negative results for the device. We found the sensor to be inaccurate, imprecise, uncomfortable, and unreliable as a system as a whole.

Possible Device Improvements

We recommend a different wireless system. Bluetooth operates on 2.4GHz which can be blocked by water, a molecule that we are full of. It might be better to use a 933MHz carrier wave and have an external receiver on the the iPhone. As the device has hard and somewhat sharp edges, it might be worth looking into a design that uses a soft rubber like material (silicone perhaps) to contain the components. This way the entire device could conform to the user's shape, making it more comfortable to wear.




LAB 3B WRITE-UP

Target Population and Need



Device Design



Inferential Statistics




Graph