Testing of RT-PCR assay to control DNA concentration for Solexa sequencing
Standard curve using serial dilution of PhiX control DNA.
Simultaneous test. Samples were tested for RT-PCR and were clustered for sequencing on the same day. Their DNA concentration and the correlation of the concentration as measured by RT-PCR was then compared to their cluster number. All DNA samples had been thought to be at 10nM.
[[Image:|400px]]
Each sample is individually colored and represented as two replicates.