The SPRI-works system  consists of three main components: the SPRI-TE Nucleic Acid Extractor, a method card that controls liquid handling, and a cartridge containing all the reagents required to prepare a single library.  
The SPRI-works system  consists of three main components: the SPRI-TE Nucleic Acid Extractor, a method card that controls liquid handling, and a cartridge containing all the reagents required to prepare a single library.  


This platform utilizes DNA binding magnetic beads to purify fragmented DNA and perform size selection, replacing the need for column purification and gel-base size selection. There are two size range options available, 100-400bp or 300-600bp. There is also the choice to perform no size selection. When complete samples will be ready for PCR Enrichment and clean-up to complete the library prep process.  
This platform utilizes AMPureXP DNA binding magnetic beads to purify fragmented DNA and perform size selection, replacing the need for column purification and gel-based size selection. There are two size range options available, 100-400bp or 300-600bp. There is also the choice to perform no size selection. When complete samples will be ready for PCR Enrichment and clean-up to complete the library prep process.  


For more information on the SPRI-Works system please visit www.spriworks.com
For more information on the SPRI-Works system please visit www.spriworks.com

Revision as of 12:59, 2 August 2010

HOME -- SEQUENCING -- LIBRARY PREP -- HIGH-THROUGHPUT -- COMPUTING -- OTHER TECHNOLOGY

The BMC is currently testing the Beckman SPRIworks library preparation system. This platform is a fully automated library preparation system for Illumina sequencing and supports all Illumina applications except for smallRNA-Seq. It performs all of the traditional Illumina protocol steps listed on the Illumina Library Preparation page, except for Enrichment, can prepare up to ten libraries for Illumina sequencing in 5 hours.

How it Works

The SPRI-works system consists of three main components: the SPRI-TE Nucleic Acid Extractor, a method card that controls liquid handling, and a cartridge containing all the reagents required to prepare a single library.

This platform utilizes AMPureXP DNA binding magnetic beads to purify fragmented DNA and perform size selection, replacing the need for column purification and gel-based size selection. There are two size range options available, 100-400bp or 300-600bp. There is also the choice to perform no size selection. When complete samples will be ready for PCR Enrichment and clean-up to complete the library prep process.

For more information on the SPRI-Works system please visit www.spriworks.com

Sample Submission

All samples submitted for direct SPRI-Works library preparation must be submitted with user supplied adapter mix. Please note that samples will need to be submitted separately for Illumina sequencing.