BioMicroCenter:Wafergen: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
Line 46: Line 46:
|}
|}
<br>
<br>
{|border=1 alignment="left"
|-
!Primer-sets
!Samples
!Wells per primer-set in primer-set sourceplate
!Wells per sample in sample sourceplate
!Required total volume of primer-set wells (uL)
!Volume of primer-set in each well (uL)
!Required total volume of sample wells (uL)
!Volume of sample in each well (uL)
|-
|12
|384
|4
|1
|17.9
|4.5
|11.7
|3.0
|-
|24
|216
|2
|1
|17.9
|4.5
|12.4
|3.1
|-
|36
|144
|1
|1
|20.3
|5.1
|13.2
|3.3
|-
|48
|108
|1
|1
|17.9
|4.5
|14.0
|3.5
|-
|54
|96
|1
|1
|17.1
|4.3
|14.4
|3.6
|-
|72
|72
|1
|1
|15.6
|3.9
|15.6
|3.9
|-
|80
|64
|1
|1
|15.1
|3.8
|16.1
|4.1
|-
|96
|54
|1
|1
|14.4
|3.6
|17.1
|4.3
|-
|120
|42
|1
|1
|13.7
|3.5
|18.9
|4.8
|-
|144
|36
|1
|1
|13.2
|3.3
|20.3
|5.1
|-
|216
|24
|1
|2
|12.4
|3.1
|17.9
|4.5
|-
|248
|20
|1
|2
|12.2
|3.1
|19.4
|4.9
|-
|296
|16
|1
|4
|12.0
|3.0
|16.1
|4.1
|-
|384
|12
|1
|4
|11.7
|3.0
|17.9
|4.5
|}<br>


Wafergen recommends samples normalized to a concentration of 10 ng/uL. Although normalization is not required, the minimum acceptable concentration is 1 ng/uL.<br>
Wafergen recommends samples normalized to a concentration of 10 ng/uL. Although normalization is not required, the minimum acceptable concentration is 1 ng/uL.<br>

Revision as of 11:54, 11 April 2013

HOME -- SEQUENCING -- LIBRARY PREP -- HIGH-THROUGHPUT -- COMPUTING -- OTHER TECHNOLOGY
Image of Wafergen's SmartChip

The Wafergen SmartChip System is a highly flexible nanoliter qPCR instrument. The system is based on aluminum chips that contain a grid of 72x72 nanowells. Each well can hold up to 150ul of volume. The system itself is composed of two halves: a robot for loading the chips and a PCR machine for cycling and scanning the chips.


.

REACTION SETUP

Wafergen reactions come in two flavors: prefilled arrays and "MyDesign". Prefilled arrays have specific primer sets laid out across the nanowells. MyDesign arrays are filled in two steps: cDNA followed by primers.

Primers and cDNA are filled from specific 384 well plates in master mix already diluted to 1x. Technical replicates must be pipetted the desired number of times on the plate. eg. if you want to run your primers 3x on the cDNA, you should use 3 wells for each primer set and only 1 for each cDNA. Even though each dispense uses only 50nL, more material is required for pipetting. The matrix below outlines the preprogrammed sample/primer set ratios and the volumes of sample and primer required.

Samples Primer Sets
Number Volume Number Volume
12 18ul x4 wells 384 12ul
24 18ul x2 wells 216 12.5ul
36 20.5ul 144 13.5ul
48 18ul 108 14ul
54 17.5ul 96 14.5ul
64 16.5ul 80 15.5ul
72 16ul 72 16ul
80 15.5ul 64 16.5ul
96 14.5 54 17.5ul
108 14ul 48 18ul
144 13.5ul 36 20.5ul
216 12.5ul 24 18ul x2 wells
384 12ul 12 18ul x4 wells


Wafergen recommends samples normalized to a concentration of 10 ng/uL. Although normalization is not required, the minimum acceptable concentration is 1 ng/uL.

Primers should be reconstituted to a concentration of 1.5 uM.

Also required are two Excel spreadsheets detailing the names of the samples and the primers, respectively, arranged in the order they will be dispensed.

RUNNING THE CHIPS

The Wafergen system is offered as a walkup service in the BioMicro Center. However, due to the complexity of the Wafergen system, a single brief training session is not sufficient for us to be comfortable letting users operate on their own. Instead, new users must use the robot under the guidance of the BioMicro Center technician for the first 2-3 runs (there is a higher cost associated with this). Once the user has completed their guided usage, you are free to use the system without supervision.

Users may sign up for the Wafergen system on the calendar‎. New users should confirm their schedule with BioMicro Center staff (biomicro@mit.edu).

DATA ANALYSIS


PRICING

Each chip currently costs $150.
Training runs will cost an additional $100 for the first three runs.
384 plates specific for the Wafergen system (two required) are $10 each.