| == SPEX==
| |
|
| |
| 1. Turn on the lamp and wait for 30 min for the lamp to warm up.
| |
|
| |
| 2. Launch the software: Instrument control center
| |
|
| |
| 3. Temperature set up
| |
| Click the third square icon “Visual setup”
| |
| Temperature control icon: Turn on and Set to 25 degreeC
| |
| Choose T bath (internal).
| |
|
| |
| 4. Wash the cuvettes.
| |
| Distilled Water (DW) (10 times) 70% ethanol (twice) DW (5 times) 70% ethanol (once)
| |
| Air dry the cuvettes for 1 h.
| |
|
| |
| 5. Clean the cuvettes using lens paper.
| |
|
| |
| 6. Click second square icon “real time display”.
| |
| High voltage: HV on, S 950, T 950
| |
| Bandpass : 2
| |
|
| |
| 7. Click the last icon “run CWA”.
| |
| Open the file “ROX_CWA”
| |
| Optional mode: kinetics, Integration time: 10s
| |
| Define: Repeat acquisition: 1 min (minimum) for a total of ~ min, depending on how long you would like to run the experiment.
| |
|
| |
| 8. Mixing: Mix the sample with 1X TE/Mg++ buffer inside cuvettes by pipetting for 30 times.
| |
|
| |
| 9. Start acquisition. Run sample.
| |
| The data should not exceed 1 million (1000,000).
| |
|
| |
| 10. To add more samples (for example, walker triggers) in the middle of SPEX experiment
| |
| When the status is “waiting”, click “stop”.
| |
| Add samples and mix. Replace cuvettes into the SPEX machine.
| |
|
| |
| 11. Save the data
| |
| When the status is “waiting”, click “stop”.
| |
| Click “yes”
| |
| Save data in two forms (DWA datafile, txt file)
| |
| Check if the data is properly saved.
| |
| *If you want to see the reference signals, go to options: view option and display raw R.
| |