Cfrench:anthrone

From OpenWetWare

(Difference between revisions)
Jump to: navigation, search
(Assaying polysaccharides with the anthrone assay)
Line 3: Line 3:
'''Equipment and Instrument'''
'''Equipment and Instrument'''
-
1. 19×150mm glass test tube
+
*1. 19×150mm glass test tube
-
2. Cuvettes
+
*2. Cuvettes
-
3. Glass pipette and pipette aid
+
*3. Glass pipette and pipette aid
-
4. Spectrometer
+
*4. Spectrometer
-
5. Chemical hood
+
*5. Chemical hood
Line 17: Line 17:
'''Preparation of Reagent and Standard Solution'''
'''Preparation of Reagent and Standard Solution'''
-
1. 2% anthrone in ethyl acetate ['''SAFETY NOTES''': Ethyl acetate is highly flammable in presence of open flames and sparks, of heat. Be careful when using it.]
+
*1. 2% anthrone in ethyl acetate ['''SAFETY NOTES''': Ethyl acetate is highly flammable in presence of open flames and sparks, of heat. Be careful when using it.]
-
2. Glucose standard solution (0~80µg):
+
*2. Glucose standard solution (0~80µg):
-
<nowiki>  </nowiki> SolA: 1mg/2mL std. (100mL): 50mg glucose + 100mL H2O (50mg glucose in total)
+
**SolA: 1mg/2mL std. (100mL): 50mg glucose + 100mL H2O (50mg glucose in total)
    
    
-
  SolB: 100µg/2mL std. (50mL): 5mL SolA + 45mL H2O (2.5mg glucose in total)
+
**SolB: 100µg/2mL std. (50mL): 5mL SolA + 45mL H2O (2.5mg glucose in total)
   1) 80µg/2mL std. (10mL): 8mL SolB + 2mL H2O (400µg glucose in total)
   1) 80µg/2mL std. (10mL): 8mL SolB + 2mL H2O (400µg glucose in total)

Revision as of 18:35, 23 August 2008

Assaying polysaccharides with the anthrone assay

Equipment and Instrument

  • 1. 19×150mm glass test tube
  • 2. Cuvettes
  • 3. Glass pipette and pipette aid
  • 4. Spectrometer
  • 5. Chemical hood


Preparation of Reagent and Standard Solution

  • 1. 2% anthrone in ethyl acetate [SAFETY NOTES: Ethyl acetate is highly flammable in presence of open flames and sparks, of heat. Be careful when using it.]
  • 2. Glucose standard solution (0~80µg):
    • SolA: 1mg/2mL std. (100mL): 50mg glucose + 100mL H2O (50mg glucose in total)
    • SolB: 100µg/2mL std. (50mL): 5mL SolA + 45mL H2O (2.5mg glucose in total)
  1) 80µg/2mL std. (10mL): 8mL SolB + 2mL H2O (400µg glucose in total)
  2) 50µg/2mL std. (20mL): 10mL SolB + 10mL H2O (500µg glucose in total)
  3) 25µg/2mL std. (20mL): 5mL SolB + 15mL H2O (250µg glucose in total)
  4) 10µg/2mL std. (20mL): 2mL SolB + 18mL H2O (100µg glucose in total)
  5) 5µg/2mL std, (20mL): 1mL SolB + 19mL H2O (50µg glucose in total)