Endy:Gel imager

From OpenWetWare

(Difference between revisions)
Jump to: navigation, search
Current revision (17:19, 4 September 2007) (view source)
(Other Information)
 
Line 16: Line 16:
==Other Information==
==Other Information==
-
[http://www.syngene.com/cg2short.asp Syngene site]
+
[http://www.syngene.com/ Syngene site]

Current revision

Contents

Syngene ChemiGenius 2XE Imager

More info to come.

Contact

Sri Kosuri

Address

smb://sblab5.mit.edu


Problems/Fixes

In general the machine has been very good to us through high usage. Here are some comments with things going wrong.

  1. The computer that came with the machine had a motherboard failure of sometype. We were sent a new one. It has been working great since.
  2. The white light table initially had problems with fraying of the wire feeding it power. This was occuring because the wire was getting caught when we were pulling out the table. We fixed it by splicing in a new cable (sent to us by syngene), therefore extending the power cord and reducing stress on the wire.

Other Information

Syngene site

Personal tools