Endy:Victor3 Absorbance and fluorescence backgrounds: Difference between revisions
Barry Canton (talk | contribs) No edit summary |
No edit summary |
||
(4 intermediate revisions by 2 users not shown) | |||
Line 32: | Line 32: | ||
|1708 | |1708 | ||
|---- | |---- | ||
| | |||
| | |||
| | |||
|---- | |||
|EZ media + 0.4% Glycerol + Kan | |||
|0.046 | |||
|833 | |||
|---- | |||
|EZ media + 0.4% Glycerol + Strep | |||
|0.055 | |||
|738 | |||
|} | |||
| | |||
[[Image:Victor3platereadermediabackground.jpg|thumb|700px|left|'''Plot of media fluorescence (measured using GFP filters) versus time''']] | |||
This plot shows the background fluorescence of [[Neidhardt EZ Rich Defined | EZ rich media]] from Teknova + 0.4% glycerol + antibiotic over time. This is data from two separate runs done on different days (circles and squares). Each run is an average of 3 replicates. | |||
This experiment was done on the Endy lab Victor3 plate reader using 96 well plates from the Endy lab but using media and antibiotic from the [[Knight Lab]]. Water was injected into the wells over the course of the experiment to counteract evaporation, according to [[Endy:Victor3 counteracting evaporation in long time courses | this protocol]]. | |||
The reason for the lower background of EZ rich media + streptomycin for one of the days is unknown. | |||
This plot is intended to demonstrate fluctuation in background (media-only) fluorescence measurements over time. This data is '''very preliminary''' and has not been independently verified. Use at your own risk! | |||
<br style="clear:both" /> | |||
[[Image:Victor3platereadermediabackgroundabsorbance.jpg|thumb|700px|left|'''Plot of media absorbance versus time''']] | |||
This plot shows the background absorbance of [[Neidhardt EZ Rich Defined | EZ rich media]] from Teknova + 0.4% glycerol + antibiotic over time. This is data from two separate runs done on different days (circles and squares). Each run is an average of 3 replicates. | |||
This experiment was done on the Endy lab Victor3 plate reader using 96 well plates from the Endy lab but using media and antibiotic from the [[Knight Lab]]. Water was injected into the wells over the course of the experiment to counteract evaporation, according to [[Endy:Victor3 counteracting evaporation in long time courses | this protocol]]. | |||
This plot is intended to demonstrate fluctuation in background (media-only) absorbance measurements over time. This data is '''very preliminary''' and has not been independently verified. Use at your own risk! | |||
<br style="clear:both" /> |
Latest revision as of 09:14, 17 March 2006
Sample | Absorbance | GFP-Separated |
---|---|---|
Empty well | 0.032 | 941 |
Water | 0.042 | 924 |
EZ media + 0.4% Glycerol | 0.038 | 1035 |
EZ media + 0.4% Glycerol + Amp | 0.04 | 1003 |
EZ media + 0.4% Glycerol + Tet | 0.047 | 943 |
EZ media + 0.4% Glycerol + Amp + Tet | 0.048 | 978 |
EZ media + 0.4% Glucose + Amp + Tet | 0.047 | 1708 |
EZ media + 0.4% Glycerol + Kan | 0.046 | 833 |
EZ media + 0.4% Glycerol + Strep | 0.055 | 738 |
This plot shows the background fluorescence of EZ rich media from Teknova + 0.4% glycerol + antibiotic over time. This is data from two separate runs done on different days (circles and squares). Each run is an average of 3 replicates.
This experiment was done on the Endy lab Victor3 plate reader using 96 well plates from the Endy lab but using media and antibiotic from the Knight Lab. Water was injected into the wells over the course of the experiment to counteract evaporation, according to this protocol.
The reason for the lower background of EZ rich media + streptomycin for one of the days is unknown.
This plot is intended to demonstrate fluctuation in background (media-only) fluorescence measurements over time. This data is very preliminary and has not been independently verified. Use at your own risk!
This plot shows the background absorbance of EZ rich media from Teknova + 0.4% glycerol + antibiotic over time. This is data from two separate runs done on different days (circles and squares). Each run is an average of 3 replicates.
This experiment was done on the Endy lab Victor3 plate reader using 96 well plates from the Endy lab but using media and antibiotic from the Knight Lab. Water was injected into the wells over the course of the experiment to counteract evaporation, according to this protocol.
This plot is intended to demonstrate fluctuation in background (media-only) absorbance measurements over time. This data is very preliminary and has not been independently verified. Use at your own risk!