Endy:Victor3 Service History

From OpenWetWare

(Difference between revisions)
Jump to: navigation, search
Line 2: Line 2:
''Error message received relating to shaking in long time courses.''
''Error message received relating to shaking in long time courses.''
*'''Error message''' - Plate holder moving error at measurement unit.  The holder cannot move freely .  The EEPROM parameters are corrupted or there is a mechanical/electronic fault is the plate holder unit.  Error number 4004.
*'''Error message''' - Plate holder moving error at measurement unit.  The holder cannot move freely .  The EEPROM parameters are corrupted or there is a mechanical/electronic fault is the plate holder unit.  Error number 4004.
-
*'''Service call -''' placed 10am 8/29/05.
+
*'''Service call -''' placed 10am 8/29/05. Confirmation number 310586013
*'''Resolution -''' ?
*'''Resolution -''' ?

Revision as of 11:47, 29 August 2005

8/29/05

Error message received relating to shaking in long time courses.

  • Error message - Plate holder moving error at measurement unit. The holder cannot move freely . The EEPROM parameters are corrupted or there is a mechanical/electronic fault is the plate holder unit. Error number 4004.
  • Service call - placed 10am 8/29/05. Confirmation number 310586013
  • Resolution - ?