Background

96 well plates with transparent bases cost ~$4 each. It was unclear if washing plates (bleach, followed by ethanol) led to greater variation in readings.

Method

The GFP and absorbance of every well on a new and a washed plate were measured. The wells were empty.

Results

Histograms of the GFP readings for each plate are shown below.

New plate
New plate
Old plate
Old plate

\br

Conclusion

  • The washed plate results in a higher average background reading.
  • The washed plate actually results in tighter distribution of readings.
  • For most applications, washed plates are perfectly acceptable.