Glow Research Auto Glow
Operation Procedure
Start
- Turn on the vacuum pump at the pump
- Press AC ON
- Insert sample in chamber
- Place chamber lid
- Close the door properly (The RF cannot be activated unless the door is closed)
- Press Vacuum
- Press Tuning Auto/Man (Light turns on)
- Set the timer as desired (30 s recommended)
- Turn on gas flow
- Press Process Start; the process will begin automatically once the pressure falls below 0.5 Torr.
- You may press Abort Process if you want to terminate manually
Stop
- Push Vacuum to vent to atmosphere
- Stop flow
- Remove sample
- Turn off the machine
- Turn off the vacuum pump
Calibration Procedure
- Turn on the vacuum pump at the pump
- Press AC ON
- Press Vacuum
- Once the pressure falls to to 0.5 Torr, press Process Start
- Turn on gas flow
- Switch Selector to RF Set Point
- Set RF to Zero
- Switch Selector to display pressure
- Turn on Gas flow
- Set the pressure to 1 Torr by adjusting the gas flow
- Press Abort
- Press Vacuum
- Close Gas Flow
- Adjust the “Power set point” to 150 watts