==Calibration Procedure==
==Calibration Procedure==
#Turn on the vacuum pump at the pump
#Turn on the vacuum pump at the pump

Revision as of 15:14, 5 April 2010

Auto Glow

Glow Research Auto Glow

Operation Procedure

Start

  1. Turn on the vacuum pump at the pump
  2. Press AC ON
  3. Insert sample in chamber
  4. Place chamber lid
  5. Close the door properly (The RF cannot be activated unless the door is closed)
  6. Press Vacuum
  7. Press Tuning Auto/Man (Light turns on)
  8. Set the timer as desired (30 s recommended)
  9. Turn on gas flow
  10. Press Process Start; the process will begin automatically once the pressure falls below 0.5 Torr.
  11. You may press Abort Process if you want to terminate manually

Stop

  1. Push Vacuum to vent to atmosphere
  2. Stop flow
  3. Remove sample
  4. Turn off the machine
  5. Turn off the vacuum pump


Calibration Procedure

  1. Turn on the vacuum pump at the pump
  2. Press AC ON
  3. Press Vacuum
  4. Once the pressure falls to to 0.5 Torr, press Process Start
  5. Turn on gas flow
  6. Switch Selector to RF Set Point
  7. Set RF to Zero
  8. Switch Selector to display pressure
  9. Turn on Gas flow
  10. Set the pressure to 1 Torr by adjusting the gas flow
  11. Press Abort
  12. Press Vacuum
  13. Close Gas Flow
  14. Adjust the “Power set point” to 150 watts