Godin:Protocols/Glow Research Auto Glow
From OpenWetWare
< Godin:Protocols(Difference between revisions)
Current revision (12:03, 27 March 2012) (view source) |
|||
| Line 7: | Line 7: | ||
You can find an instructional video of the steps described here created by GlowResearch at: http://www.youtube.com/user/glowresearchplasma#p/a/u/1/tE8pfiYQxr4 | You can find an instructional video of the steps described here created by GlowResearch at: http://www.youtube.com/user/glowresearchplasma#p/a/u/1/tE8pfiYQxr4 | ||
| - | == | + | ==Start== |
| - | #Turn on the vacuum pump by turning on the power bar | + | #Turn on the vacuum pump by turning on the power bar and open the valve. |
| - | #Press ''AC | + | #Press ''AC ON''. |
| + | #Switch Selector to ''RF Set Point''. | ||
| + | #Set ''RF Set Point'' to zero. | ||
| + | #Switch ''Selector'' to ''Pressure''. | ||
| + | #Set timer to 5 min. | ||
#Make sure the chamber is clean. Use ethanol wipes as necessary. | #Make sure the chamber is clean. Use ethanol wipes as necessary. | ||
| - | # | + | #Place sample in chamber. |
| - | + | #Press ''Vacuum''. You may have to hold the door in place to ensure a proper seal. Make sure the door is well centered, a misaligned door will cause the chamber to leak. | |
| - | + | #Ensure the ''Tuning Auto/Man'' button is not pressed (light is off). | |
| - | + | #Once the pressure falls below 0.5 Torr, press ''Process Start''. | |
| - | #Press ''Vacuum'' | + | |
| - | #Ensure the ''Tuning Auto/Man | + | |
| - | #Once the pressure falls below 0.5 Torr, press ''Process Start'' | + | |
#Turn on gas line by flipping GAS 1 toggle switch up; Note that GAS 2 is connected to the outside of the cleanroom by a tube, for use with argon. | #Turn on gas line by flipping GAS 1 toggle switch up; Note that GAS 2 is connected to the outside of the cleanroom by a tube, for use with argon. | ||
| - | #Let the pressure | + | #Let the pressure go down until 0.85 torr. |
| - | # | + | #Turn off the gas line. |
| - | #Press ''Process | + | #Press ''Process Abort''. |
| - | #Turn | + | #Switch ''Selector'' to ''RF Set Point''. |
| - | + | #Adjust the RF Set Point rotating the knob (30W recommended). | |
| - | + | #Press ''Tuning Auto/Man'' (light turns on). | |
| - | + | #Rotate Selector to Pressure. | |
| + | #Set the timer as desired (30s recommended). Press ''Process Abort'' to reset timer. | ||
| + | #Press ''Process Start'', once the pressure reading is below 0.5 Torr. Your process will begin. | ||
| + | #Turn gas line on '''IMMEDIATELY''' by flipping GAS 1 toggle switch up; You may press ''Abort Process'' if you want to terminate manually. | ||
<br> | <br> | ||
| + | '''IMPORTANT:''' Make sure the pressure stabilizes between 0.8-1.3 Torr during your process. Multiple “beeps” indicate you’re operating outside this range. If pressure is incorrect, immediately abort, and make sure you went through all the steps. <br> | ||
| + | '''ALSO IMPORTANT:''' Never, ever rotate any knobs on the black box. System flow rate has been carefully calibrated. Only flipping toggle switches up and down is allowed. | ||
| - | + | ==Stop== | |
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| - | + | ||
| + | #Stop gas flow. | ||
| + | #Press ''Vacuum'' to vent to atmosphere. Make sure GAS 3 toggle switch is up, as chamber now vents through the black box. You can leave it open at all times. | ||
| + | #Remove sample. | ||
| + | #Make sure the chamber is clean. Use ethanol wipes as necessary. | ||
| + | #Turn off the machine. | ||
| + | #Close the valve. | ||
| + | #Turn off the vacuum pump (power bar). | ||
<br> | <br> | ||
| + | '''Having problems with the system? Talk to Jason or Dominique. They will help you for troubleshooting.''' | ||