Koch Lab:Protocols/Photolithography: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
No edit summary
 
(15 intermediate revisions by 2 users not shown)
Line 1: Line 1:
Legend: "->" denotes physically pushing the button on the machine with the text that appears directly after symbol.  
=Legend=
"->" denotes physically pushing the button on the machine with the text that appears directly after symbol.  


=General Outline of process=
#Spin
#Pre bake
#Expose
#Post bake
[[Image:LopezLab Hotplate.jpg|thumb|right|300 px|Hot plate for prebaking]]
=Procedure=
Preheat hotplate for prebaking wafer
Preheat hotplate for prebaking wafer
*->set
*->set
*->plate temp
*->plate temp
*->"desired temp"
*->"Desired temp"
**->9
**->5
*->ent
*->ent
[[Image:LopezLab Spincoater.jpg|thumb|right|300 px|Spin coater]]
[[Image:LopezLab Compressed Nitrogen.jpg|thumb|right|300 px|Nitrogen knob]]
[[Image:LopezLab UV Exposer.jpg|thumb|right|300 px|UV exposer]]
[[Image:IMG 0020.jpg|thumb|right|300 px|Developing wafer]]


Applying photo resist
Applying photo resist
Line 19: Line 35:
#Turn on vacuum pump (manually)
#Turn on vacuum pump (manually)
#*->center
#*->center
#Manually adjust wafer if needed (repeat previous 2 steps)
#Manually adjust wafer if needed (repeat previous 2 steps until wafer is centered sufficiently)
#poor photo resist
#Poor photo resist on the top side of the wafer
#*->spin
#*->spin
#*->ok
#*->ok
#**NOTE: bubbles on wafer are normal
#Move wafer to hot plate
#Move wafer to hot plate
#*wait 15 minutes
#*wait 15 minutes
Line 37: Line 54:
#*Load wafer on tray below
#*Load wafer on tray below
#**->chuck vac
#**->chuck vac
#Push slide in
#*Push slide in
#**->mask load
#**->mask load
#**->visual align
#**->visual align
Line 43: Line 60:
#**->cont (pushes wafer against mask)
#**->cont (pushes wafer against mask)
#*Set time
#*Set time
#**->manual espose
#**->manual expose
#*->chuck vac
#*->chuck vac
#*remove wafer
#*remove wafer
#Post bake (5 mins)
#*->mask load
#*->mask clamp
#Take off slide
#Turn everything off!!!
#Place wafer in glass dish
#*NOTE: It is imperative that glass is used, if a petri dish is used the developer with fuse the wafer to the dish.
#Poor in developer (6 mins)
#Take wafer out of developer
#*At this point you can spray the wafer with ethanol to remove residual developer
#*Can also use Nitrogen gas to blow off fluid/debris from wafer
#Dispose of developer in designated bottle under fume hood.

Latest revision as of 15:30, 21 October 2009

Legend

"->" denotes physically pushing the button on the machine with the text that appears directly after symbol.

General Outline of process

  1. Spin
  2. Pre bake
  3. Expose
  4. Post bake
Hot plate for prebaking

Procedure

Preheat hotplate for prebaking wafer

  • ->set
  • ->plate temp
  • ->"Desired temp"
    • ->9
    • ->5
  • ->ent
Spin coater
Nitrogen knob
UV exposer
Developing wafer

Applying photo resist

  1. Place wafer on spin coating machine
    • ->speed (input speed)
    • ->accel (input acceleration)
    • ->time (input time)
    • ->save->back
    • ->run process
    • ->load recipe
    • ->"file name"
  2. Manually center wafer by "eyeballing" it (this may take several tries)
  3. Turn on vacuum pump (manually)
    • ->center
  4. Manually adjust wafer if needed (repeat previous 2 steps until wafer is centered sufficiently)
  5. Poor photo resist on the top side of the wafer
    • ->spin
    • ->ok
      • NOTE: bubbles on wafer are normal
  6. Move wafer to hot plate
    • wait 15 minutes
  7. Hook up vacuum hose to UV exposer
  8. Turn on Nitrogen
  9. UV exposer
    • Pull white lever
    • ->start
    • wait for 200W
    • ->Power
    • ->visual align
    • Tape mask onto glass (place glass side up)
      • ->mask clamp
    • Load wafer on tray below
      • ->chuck vac
    • Push slide in
      • ->mask load
      • ->visual align
      • ->wafer load
      • ->cont (pushes wafer against mask)
    • Set time
      • ->manual expose
    • ->chuck vac
    • remove wafer
  10. Post bake (5 mins)
    • ->mask load
    • ->mask clamp
  11. Take off slide
  12. Turn everything off!!!
  13. Place wafer in glass dish
    • NOTE: It is imperative that glass is used, if a petri dish is used the developer with fuse the wafer to the dish.
  14. Poor in developer (6 mins)
  15. Take wafer out of developer
    • At this point you can spray the wafer with ethanol to remove residual developer
    • Can also use Nitrogen gas to blow off fluid/debris from wafer
  16. Dispose of developer in designated bottle under fume hood.