"->" denotes physically pushing the button on the machine with the text that appears directly after symbol. | |||
=General Outline of process= | |||
#Spin | |||
#Pre bake | |||
#Expose | |||
#Post bake | |||
[[Image:LopezLab Hotplate.jpg|thumb|right|300 px|Hot plate for prebaking]] | |||
=Procedure= | |||
Preheat hotplate for prebaking wafer | Preheat hotplate for prebaking wafer | ||
*->set | *->set | ||
*->plate temp | *->plate temp | ||
*->" | *->"Desired temp" | ||
**->9 | |||
**->5 | |||
*->ent | *->ent | ||
[[Image:LopezLab Spincoater.jpg|thumb|right|300 px|Spin coater]] | |||
[[Image:LopezLab Compressed Nitrogen.jpg|thumb|right|300 px|Nitrogen knob]] | |||
[[Image:LopezLab UV Exposer.jpg|thumb|right|300 px|UV exposer]] | |||
[[Image:IMG 0020.jpg|thumb|right|300 px|Developing wafer]] | |||
Applying photo resist | Applying photo resist | ||
Line 19: | Line 35: | ||
#Turn on vacuum pump (manually) | #Turn on vacuum pump (manually) | ||
#*->center | #*->center | ||
#Manually adjust wafer if needed (repeat previous 2 steps) | #Manually adjust wafer if needed (repeat previous 2 steps until wafer is centered sufficiently) | ||
# | #Poor photo resist on the top side of the wafer | ||
#*->spin | #*->spin | ||
#*->ok | #*->ok | ||
#**NOTE: bubbles on wafer are normal | |||
#Move wafer to hot plate | #Move wafer to hot plate | ||
#*wait 15 minutes | #*wait 15 minutes | ||
Line 37: | Line 54: | ||
#*Load wafer on tray below | #*Load wafer on tray below | ||
#**->chuck vac | #**->chuck vac | ||
#Push slide in | #*Push slide in | ||
#**->mask load | #**->mask load | ||
#**->visual align | #**->visual align | ||
Line 43: | Line 60: | ||
#**->cont (pushes wafer against mask) | #**->cont (pushes wafer against mask) | ||
#*Set time | #*Set time | ||
#**->manual | #**->manual expose | ||
#*->chuck vac | #*->chuck vac | ||
#*remove wafer | #*remove wafer | ||
#Post bake (5 mins) | |||
#*->mask load | |||
#*->mask clamp | |||
#Take off slide | |||
#Turn everything off!!! | |||
#Place wafer in glass dish | |||
#*NOTE: It is imperative that glass is used, if a petri dish is used the developer with fuse the wafer to the dish. | |||
#Poor in developer (6 mins) | |||
#Take wafer out of developer | |||
#*At this point you can spray the wafer with ethanol to remove residual developer | |||
#*Can also use Nitrogen gas to blow off fluid/debris from wafer | |||
#Dispose of developer in designated bottle under fume hood. |
Latest revision as of 15:30, 21 October 2009
Legend
"->" denotes physically pushing the button on the machine with the text that appears directly after symbol.
General Outline of process
- Spin
- Pre bake
- Expose
- Post bake
Procedure
Preheat hotplate for prebaking wafer
- ->set
- ->plate temp
- ->"Desired temp"
- ->9
- ->5
- ->ent
Applying photo resist
- Place wafer on spin coating machine
- ->speed (input speed)
- ->accel (input acceleration)
- ->time (input time)
- ->save->back
- ->run process
- ->load recipe
- ->"file name"
- Manually center wafer by "eyeballing" it (this may take several tries)
- Turn on vacuum pump (manually)
- ->center
- Manually adjust wafer if needed (repeat previous 2 steps until wafer is centered sufficiently)
- Poor photo resist on the top side of the wafer
- ->spin
- ->ok
- NOTE: bubbles on wafer are normal
- Move wafer to hot plate
- wait 15 minutes
- Hook up vacuum hose to UV exposer
- Turn on Nitrogen
- UV exposer
- Pull white lever
- ->start
- wait for 200W
- ->Power
- ->visual align
- Tape mask onto glass (place glass side up)
- ->mask clamp
- Load wafer on tray below
- ->chuck vac
- Push slide in
- ->mask load
- ->visual align
- ->wafer load
- ->cont (pushes wafer against mask)
- Set time
- ->manual expose
- ->chuck vac
- remove wafer
- Post bake (5 mins)
- ->mask load
- ->mask clamp
- Take off slide
- Turn everything off!!!
- Place wafer in glass dish
- NOTE: It is imperative that glass is used, if a petri dish is used the developer with fuse the wafer to the dish.
- Poor in developer (6 mins)
- Take wafer out of developer
- At this point you can spray the wafer with ethanol to remove residual developer
- Can also use Nitrogen gas to blow off fluid/debris from wafer
- Dispose of developer in designated bottle under fume hood.