Lidstrom Lab French Press: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
No edit summary
Line 6: Line 6:
* 4- Cylinder cap
* 4- Cylinder cap
* 5- Spigot
* 5- Spigot
[[Image:FPparts.jpg|600px]]
[[Image:numberedFP.jpg|600px]]


Prepare the equipment:
Prepare the equipment:

Revision as of 09:45, 14 June 2013

French Press

Collect your equipment:

  • 1- Cell cylinder
  • 2- Screw valve
  • 3- Piston
  • 4- Cylinder cap
  • 5- Spigot

Prepare the equipment:

  • Check the condition of the O-rings on the piston and cylinder bottom before starting your extraction
- If they are shredded, sticky or warped you should change them
- There are 2 types of O-ring, flat and rounded; the rounded O-ring should be positioned on the outside
  • Check the condition of the nylon ball housed in the screw valve
- The ball must be rounded to effectively adjust pressure
- Flattened or squared off balls should be changed out
- Use a razor blade to pry the malformed ball out of the housing and place a new ball in the opening
  • Refridgerate setup at 4°C until thoroughly cold

You must have someone show you how to operate the press itself if you have never used it, but here is an overview of the press controls:

  • 1- Pressure gauge
  • 2- Pressure dial (increase/decrease psi)
  • 3- Ratio selector (up/down)