|
|
Line 1: |
Line 1: |
| [[NikonEclipseTI| Nikon Eclipse TI]] | | [[NikonEclipseTI| Nikon Eclipse TI]] |
| | | [[Freezer -80| Freezer -80]] |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| Add here a list of equipment and maintenance logs for that equipment + reminders, etc.
| |
| | |
| | |
| Note on the Thermo -80C Freezer from Dipen:
| |
| | |
| Hi Megan, Jean
| |
| | |
| The technician from Thermo had a look at the freezer door today. The door was apparently getting locked up because the sensor that allows the equilibration of temperature after the door is closed, was frozen over.
| |
| We should be able to take care of this problem in the future; there is a hole in the inside part of the door that needs to be periodically scraped out, particularly in the humid summer months.
| |
Revision as of 16:12, 23 October 2012