SET-UP | |||
# Rinse out Buffer container and fill to line with YSI 2357 Buffer solution | |||
# Hit [Menu} then {2-MeasParameter] then [2-CalMethod] then [1-One Station] | |||
# Hit [Enter] twice to return to Measurement Parameter Set up Screen | |||
# Hit [3-Black] | |||
# Make sure Current BLACK probe chem is set to NONE. If it isn't, hit [1-Next chemistry] until NONE appears and hit [ENTER] to save | |||
# Hit [4-White] and set Current WHITE probe chem to Dextrose | |||
# Hit [Enter] 4 times to return to Measurement Parameter Set up Screen | |||
# Hit [Menu] to return to home screen | |||
RUN SAMPLES | |||
# Hit [Run] and wait for system to complete calibration | |||
# When Ready screen appears, hit [Sample] | |||
# Present sample to needle in Station S3 without touching needle to tube | |||
# Hit [Sample] and wait for needle to retract | |||
* Results will print out at the completion of run | |||
* To continue testing samples, hit [Sample] when Ready screen appears | |||
*Note: After 5-7 runs, system will automatically calibrate. | |||
==Notes== | ==Notes== | ||
==References== | ==References== |
Revision as of 08:57, 5 July 2011
Introduction
Method
SET-UP
- Rinse out Buffer container and fill to line with YSI 2357 Buffer solution
- Hit [Menu} then {2-MeasParameter] then [2-CalMethod] then [1-One Station]
- Hit [Enter] twice to return to Measurement Parameter Set up Screen
- Hit [3-Black]
- Make sure Current BLACK probe chem is set to NONE. If it isn't, hit [1-Next chemistry] until NONE appears and hit [ENTER] to save
- Hit [4-White] and set Current WHITE probe chem to Dextrose
- Hit [Enter] 4 times to return to Measurement Parameter Set up Screen
- Hit [Menu] to return to home screen
RUN SAMPLES
- Hit [Run] and wait for system to complete calibration
- When Ready screen appears, hit [Sample]
- Present sample to needle in Station S3 without touching needle to tube
- Hit [Sample] and wait for needle to retract
- Results will print out at the completion of run
- To continue testing samples, hit [Sample] when Ready screen appears
- Note: After 5-7 runs, system will automatically calibrate.