Smolke:Protocols/Glass Beads

From OpenWetWare

< Smolke:Protocols
Revision as of 17:07, 20 October 2009 by Josh K. Michener (Talk | contribs)
(diff) ←Older revision | Current revision (diff) | Newer revision→ (diff)
Jump to: navigation, search

Home        Contact        Internal        Protocols        Lab Members        Publications        Research       


Procedure for recycling used glass beads.


  • Lab dispenser soap
  • 95% Ethanol


  1. Collect beads from plastic containers around the benches.
  2. Soak the beads in a larger container with water.
  3. Add soap and mix well.
    • This step is to remove the excess material stuck to the beads.
  4. Rinse the beads with water to remove all trace of soap.
    • It helps to use a strainer.
  5. Rinse the beads with the Ethanol.
    • This step removes residual water.
  6. Transfer beads to glass bead jars.
    • Cover loosely with lids and allow the beads to dry (overnight). autoclaving while still wet may cause the beads to develop a filmy coating.
  7. Autoclave on gravity cyle, tighten lids on after cooling and store.