Smolke:Protocols/Guidelines

From OpenWetWare
Jump to navigationJump to search

Start Up

Instructions for machine start up and various other functions are on the lamenated guides near the VYB.

If machine is off:

  • Turn on machine by touching to screen.
  • Logon and put the machine into data acquisition mode.
  • Run back flush (~16 min).
  • Run clean with 1% hypochlorite (~10 min).
    • NOTE: 100% bleach = 15% hypochlorite.
    • NOTE: Clean only runs if machine is in the single tube mode.
  • Run calibration beads.

If machine is already on, assume it is not clean:

  • Run back flush (~16 min).
  • Run clean with 1% hypochlorite (~10 min).

Shut Down

  • When finished, ALWAYS make sure there is ENOUGH BUFFER and ENOUGH SPACE in the waste container. Otherwise, shut down will stall and leave sticky, dead, clog-inducing samples in the machine until someone notices.
  • Shut down the machine unless another user is signed up within a couple of hours after you finish, but check buffer and waste in case later user cancels.
  • Machine will automatically attempt to shut down ~2 hrs after its last fluidics function. However, you must still check buffer and waste before leaving the machine.

96-Well Plates

  • The VYB can only run ONE full 96-well plate if it has a FULL bottle of running buffer and an EMPTY waste bottle.
  • Make sure the running buffer bottle is full and waste bottle is empty before EACH 96-well plate.
  • NOTE: In the future, there will be larger buffer bottles & this may no longer be a problem.

Empty Buffer Bottles

  • Rinse with water.
  • Remove the label and lid.
  • Recycle (blue atrium bin).

Filtering Samples

ALWAYS FILTER YOUR SAMPLES.

  • Nylon filter sheets (http://www.amazon.com/Microns-Square-Opening-Thread-Diameter/dp/B000FN0OG2/ref=pd_sbs_indust_5)
    • Available in zip-lock bag above VYB bench.
    • Clean-up:
      • Mammalian samples: Soak in 10% bleach for about 1 day, then soak in water about 1 day, then dry and return sheets to bag.
      • Yeast samples: Soak in water for a couple of days, then dry and return sheets to bag. (YHW: I usually spread ethanol on it while drying)
  • Cell strainers (BD Falcon Cat. No.: 352340)
    • Not consistently stocked. Order if needed.
    • Disposable.

Quirks

  • The software is frozen.
    • If you are running a sample, leave the machine alone. It is probably busy recording a lot of data and will recover once it finishes (up to ~15 min).
  • I can't run 'Clean'.
    • Make sure the machine is set in the single tube mode.
  • My samples look weird in FlowJo.
    • In FlowJo, adjust the file properties to match the decade settings used on the VYB.