User:Abigail E. Miller/Notebook/CHEM571 Lab Project/2012/10/08
From OpenWetWare
(Difference between revisions)
(→Objective) |
(→Bench Work) |
||
| Line 10: | Line 10: | ||
==Bench Work== | ==Bench Work== | ||
| - | * | + | * 1000 ± 10μg of Au/mL of H<sub>2</sub>O stock solution |
| + | ** from Inorganic Ventures cat# AAAU1-1, lot # F2-AU04012 | ||
| + | ** the solution is certified by ICP (ion couple plasma spectroscopy) to NIST SRM 3121 | ||
| + | |||
| + | For the AAS (atomic absorption Spectrometer), the published detection limit for gold is 0.2 ppm ( a 1% absorbance)for this instrument. recent experiment is closer to 2 ppm and since 20 ppm is roughly 0.1 mM Au, Therefore I created standards ranging from 5 to 40 ppm. | ||
| + | |||
| + | * created 50 mL of each standard: 5, 8, 10, 15, 20, 25, 30, 40 ppm Au from the 1000 ppm Au stock standard. | ||
| + | ** | ||
| + | {| {{table}} | ||
| + | | align="center" |'''ppm''' | ||
| + | | align="center" |'''mL of 1000 ppm stock added''' | ||
| + | |- | ||
| + | | 5||0.25 | ||
| + | |- | ||
| + | | 8||0.40 | ||
| + | |- | ||
| + | | 10||0.50 | ||
| + | |- | ||
| + | | 15||0.75 | ||
| + | |- | ||
| + | | 20||1.00 | ||
| + | |- | ||
| + | | 25||1.25 | ||
| + | |- | ||
| + | | 30||1.50 | ||
| + | |- | ||
| + | | 40||2.00 | ||
| + | |- | ||
| + | | | ||
| + | |} | ||
==Data/Spectra== | ==Data/Spectra== | ||
Revision as of 14:49, 8 October 2012
Main project page Previous entry Next entry
| ||||||||||||||||||||
Objective
Bench Work
For the AAS (atomic absorption Spectrometer), the published detection limit for gold is 0.2 ppm ( a 1% absorbance)for this instrument. recent experiment is closer to 2 ppm and since 20 ppm is roughly 0.1 mM Au, Therefore I created standards ranging from 5 to 40 ppm.
Data/Spectra
Notes
| ||||||||||||||||||||