Objective
- To create standard solution of gold for calibration of the atomic absorption spectrometer. The students will measure the gold content of their BSA nucleated AuNPs against these standards.
Bench Work
- 1000 ± 10μg of Au/mL of solution ( 10 % v/v HCl) for the stock solution
- from Inorganic Ventures cat# AAAU1-1, lot # F2-AU04012
- the solution is certified by ICP (ion couple plasma spectroscopy) to NIST SRM 3121
For the AAS (atomic absorption Spectrometer), the published detection limit for gold is 0.2 ppm ( a 1% absorbance)for this instrument. recent experiment is closer to 2 ppm and since 20 ppm is roughly 0.1 mM Au, Therefore I created standards ranging from 5 to 40 ppm.
- created 50 mL of each standard: 5, 8, 10, 15, 20, 25, 30, 40 ppm Au
- from the 1000 ppm Au stock standard
- using 50 mL volumetric flasks
- all diluted in 1 M HCl
ppm
|
mL of 1000 ppm stock added
|
5 |
0.25
|
8 |
0.40
|
10 |
0.50
|
15 |
0.75
|
20 |
1.00
|
25 |
1.25
|
30 |
1.50
|
40 |
2.00
|
|
Data/Spectra
- N/A - solutions will be used for AAS on 10/9/2012
Notes
|