User:Beatriz Gimenez De C./Notebook/571/2014/10/14: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
No edit summary
No edit summary
Line 16: Line 16:
* Store in refrigerator overnight  
* Store in refrigerator overnight  


== Analysis of dialyzed from october 8==  
== October 8 dialysis analysis==  
* Bradford Analysis
* Bradford Analysis
** Remove 20 μL of each solution + add 200 μL of 1:3 diluted bradford (with  50mM Tris/50mM NaCl)+ further dilute to 1 mL with 50 mM Tris/NaCl
** Remove 20 μL of each solution + add 200 μL of 1:3 diluted bradford (with  50mM Tris/50mM NaCl)+ further dilute to 1 mL with 50 mM Tris/NaCl

Revision as of 18:54, 31 October 2014

Project name <html><img src="/images/9/94/Report.png" border="0" /></html> Main project page
<html><img src="/images/c/c3/Resultset_previous.png" border="0" /></html>Previous entry<html>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;</html>Next entry<html><img src="/images/5/5c/Resultset_next.png" border="0" /></html>

Tasks

SDS Page

  • Mix 10 μL 0.6 g/L lysozyme with 10μL SDS-PAGE sample buffer in 1.5 mL centrifuge vial
  • Mix 10 μL 0.12 g/L lysozyme with 10μL SDS-PAGE sample buffer in 1.5 mL centrifuge vial
  • Mix 10 μL 30:1 Au/lysozyme colloid with 10μL SDS-PAGE sample buffer in 1.5 mL centrifuge vial
  • Mix 10 μL 0.12 g/L unknown protein with 10 μL SDS-PAGE sample buffer in 1.5 mL centrifuge vial
  • Place in heating block (set at 90 °C) for 5 minutes
  • Store in refrigerator overnight

October 8 dialysis analysis

  • Bradford Analysis
    • Remove 20 μL of each solution + add 200 μL of 1:3 diluted bradford (with 50mM Tris/50mM NaCl)+ further dilute to 1 mL with 50 mM Tris/NaCl
    • Measure UV-Vis from 400 nm - 800 nm (following usual protocol: run blank of Tris/NaCl buffer + undialyzed 0.6 g/L Lysozyme solution with Bradford reagent)
  • Fluorescence of lysozyme containing solutions
    • Dilute lysozyme sample 100x for each well.
    • Run fluorescence from 300 nm -550 nm and excitation at 280 nm
  • UV-Vis
    • Run Uv-Vis on protein containing solutions obtained from dialisis
    • Run from 200 nm -400 nm
  • Ca2+ ISE
    • measured all the samples originally containing 'CaCl2
CaCl2 concentration mV measurement (mV)
5μM 31.8
50μM 40.6
500μM 40.1
5mM 56.6
50mM 78.5

Prepare 50 mL 66 mM Potassium Phosphate Buffer

  • Added 0.361 g of monobasic potassium phosphate + 0.105 g of dibasic potassium phosphate to 50 mL volumetric flask + filled with HPLC water
  • Final pH: 6.319

Prepare New Dialysis for 30:1 Colloid vs CaCl2

  • Use 3500 MWCO tubing
  • Add 30:1 colloid solution to 5 cells
  • Add 5 μM, 50 μM, 500 μM, 5 mM, and 50 mM CaCl2 to the opposite cells
  • Follow usual protocol

Prepare Calibration curve for Lysozyme

  • Bradford analysis for undialysed lysozyme solutions with concentrations 0.12, 0.3, 0.6 and 1 g/L, as described above.