* Transfer 50 μl of each solution to the Costar 96 clear bottom black side.
+
* Transfer 50 μl of each solution and a control to the Costar 96 clear bottom black side.
* Here is the software setting:
* Here is the software setting:
** Fluorescence  
** Fluorescence  
Line 28: Line 28:
** Read Height: 9.5 mm
** Read Height: 9.5 mm
* Click OK and then Run the experiment.
* Click OK and then Run the experiment.
 +
* Export the data to an excel file.
 +
* The software will deduce the control data from the sample data
'''Result'''
'''Result'''
Line 33: Line 35:
{| border="1"
{| border="1"
-
! Header text !! Header text
+
! Dilution !! RFP (AU)
|-
|-
|format modifier (not displayed) | 100% ||43125
|format modifier (not displayed) | 100% ||43125

Current revision

PcTF Subcloning in E. coli Main project page
Previous entry      Next entry

18/12/2012

RFP Plate Reader Serial Dilution

  1. Transfer 100 μl of RFP from stock to 100μl DI water. (50% solution)
  2. Transfer 100 μl of 50% solution to 100 μl DI water. (25% solution)
  3. Transfer 100 μl of 25% solution to 100 μl DI water. (12.5% solution)
  4. Transfer 100 μl of 12.5% solution to 100 μl DI water. (6.25% solution)

Plate Reader

  • Transfer 50 μl of each solution and a control to the Costar 96 clear bottom black side.
  • Here is the software setting:
    • Fluorescence
    • Endpoint
    • Full Plate
    • Excitation: 540, Emission: 600
    • Optics: Top
    • Gain: AutoScale
    • Light Source: Xenon Flash, Lamp Energy: High
    • Read Speed: Normal, Delay: 100 msec, Measurements/Data Point: 10
    • Read Height: 9.5 mm
  • Click OK and then Run the experiment.
  • Export the data to an excel file.
  • The software will deduce the control data from the sample data

Result


Dilution RFP (AU)
100% 43125
50% 16143
25% 7978
12.5% 4124
6.25% 2718