===Notes===
===Notes===
From the calibration graph for the Oligo D, there is a concentration drop-off between 50 and 100pM.  More samples between this interval could be made to determine at what exact concentration the limit occurs.  Otherwise, the trend between concentration of Oligo D and 1/<N> value is linear. The graphs were consistent with the exception of the 50pM sample, which could be due to the concentration drop-off observed in the calibration graph.<br.>
From the calibration graph for the Oligo D, there is a concentration drop-off between 50 and 100pM.  More samples between this interval could be made to determine at what exact concentration the limit occurs.  Otherwise, the trend between concentration of Oligo D and 1/<N> value is linear. The graphs were consistent with the exception of the 50pM sample, which could be due to the concentration drop-off observed in the calibration graph.<br.>
From the calibration graph for the DNA/MB samples, a concentration limit was not reached.  This means samples can be further diluted and run.  The graphs were consistent with the exception of the final sample, 100pM DNA/120pM MB, which was expected to have a higher y-intercept value. This may be a concentration limit similar to the Oligo D, but was not evident in the <N> value.  Samples will be made and run again tomorrow.<br.>
From the calibration graph for the DNA/MB samples, a concentration limit was not reached and the trend was linear.  This means samples can be further diluted and run.  The graphs were consistent with the exception of the final sample, 100pM DNA/120pM MB, which was expected to have a higher y-intercept value. This may be a concentration limit similar to the Oligo D, but was not evident in the <N> value.  Samples will be made and run again tomorrow.<br.>





Revision as of 12:38, 25 June 2013

FCS Data <html><img src="/images/9/94/Report.png" border="0" /></html> Main project page
<html><img src="/images/c/c3/Resultset_previous.png" border="0" /></html>Previous entry<html>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;</html>Next entry<html><img src="/images/5/5c/Resultset_next.png" border="0" /></html>

June 25, 2013

FCS Measurements

  • Samples prepared to be taken up to run on the FCS:
    • Antibodies with Alexa 532
    • Green fluorescent beads diluted 100x
    • Green fluorescent beads diluted 1000x
    • 100nM DNA/120nM MB
    • 1nM DNA/1.2nM MB
    • 500pM DNA/600pM MB
    • 100pM DNA/120pM MB
    • 500pM Oligo D fluorescent dye
    • 100pM Oligo D fluorescent dye

Results

Oligo D Fluorescent Dye<br.> <br.> <br.> Calibration Graph<br.> <br.> <br.> Varied Concentrations of DNA and MB<br.> <br.> <br.> Calibration Graph<br.> <br.>

Notes

From the calibration graph for the Oligo D, there is a concentration drop-off between 50 and 100pM. More samples between this interval could be made to determine at what exact concentration the limit occurs. Otherwise, the trend between concentration of Oligo D and 1/<N> value is linear. The graphs were consistent with the exception of the 50pM sample, which could be due to the concentration drop-off observed in the calibration graph.<br.> From the calibration graph for the DNA/MB samples, a concentration limit was not reached and the trend was linear. This means samples can be further diluted and run. The graphs were consistent with the exception of the final sample, 100pM DNA/120pM MB, which was expected to have a higher y-intercept value. This may be a concentration limit similar to the Oligo D, but was not evident in the <N> value. Samples will be made and run again tomorrow.<br.>