User:Matt Hartings/Notebook/AU Biomaterials Design Lab/2012/11/28: Difference between revisions
From OpenWetWare
Line 14: | Line 14: | ||
==Description== | ==Description== | ||
# | Using the Shimadzu AA-6200 in order to measure the gold content of the BSA-AuNP samples made yesterday. I made 4 standard solutions using our 1,000 μg/mL Au standard. | ||
# 4 μg/mL (0.100 mL standard in 25 mL of 1N HCl) | |||
# 8 μg/mL (0.200 mL standard in 25 mL of 1N HCl) | |||
# 10 μg/mL (0.250 mL standard in 25 mL of 1N HCl) | |||
# 20 μg/mL (0.500 mL standard in 25 mL of 1N HCl) | |||
==Data== | ==Data== |
Revision as of 08:25, 4 December 2012
Biomaterials Design Lab | <html><img src="/images/9/94/Report.png" border="0" /></html> Main project page <html><img src="/images/c/c3/Resultset_previous.png" border="0" /></html>Previous entry<html> </html>Next entry<html><img src="/images/5/5c/Resultset_next.png" border="0" /></html> |
ObjectiveMeasure AA of the BSA-AuNP samples made yesterday DescriptionUsing the Shimadzu AA-6200 in order to measure the gold content of the BSA-AuNP samples made yesterday. I made 4 standard solutions using our 1,000 μg/mL Au standard.
Data
NotesThis area is for any observations or conclusions that you would like to note.
|