<u>UV-Vis Data - uncentrifuged</u>
 +
 
 +
Taking the 70, 150, and 190 samples that I made [[User:Matt_Hartings/Notebook/AU_Biomaterials_Design_Lab/2013/12/05 | here]], I ran UV Vis. I noted that the tube for the 150 sample was tinged a slight purple color, indicating that some AuNPs had adsorbed to its surface. I expect a slight diminishing of the signal for both UVVis and AA for this sample.
 +
 
 +
 
 +
<u>AA - uncentrifuged</u>
 +
 
 +
Using the same samples as above ... I also made 6 standard concentrations for a calibration curve
 +
# 0
 +
## 10mL of water
 +
# 10
 +
## 100uL of 1000±10ug/mL Au standard into a 10mL volumetric flask with water
 +
# 20
 +
## 200uL of 1000±10ug/mL Au standard into a 10mL volumetric flask with water
 +
# 30
 +
## 300uL of 1000±10ug/mL Au standard into a 10mL volumetric flask with water
 +
# 40
 +
## 400uL of 1000±10ug/mL Au standard into a 10mL volumetric flask with water
 +
# 50
 +
## 500uL of 1000±10ug/mL Au standard into a 10mL volumetric flask with water
 +
 
 +
Here is the data for these samples
 +
 
 +
[[Image:20140310 HRPAuNP AAdata.png|400px]]
 +
 
 +
<u>UVVis - centrifuged</u>
 +
 
 +
I centrifuged 1mL of each sample in the centrifuge in 207 for 30 minutes at 13200rpm.
==Notes==
==Notes==

Revision as of 14:19, 10 March 2014

Biomaterials Design Lab Main project page
Previous entry      Next entry



Objective

I was looking at the UVVis data for HRP AuNPs today. I noticed that the student run data didn't qualitatively match with what the solutions looked like. The students all centrifuged their samples (30 min max speed) before taking their data (both AA and UV Vis). I wanted to take AA and UV Vis data for uncentrifuged samples. I also took data of centrifuged samples (with the remaining sample after AA).

Description

  1. Add experimental record here. Include what, how, and why...

Data

UV-Vis Data - uncentrifuged

Taking the 70, 150, and 190 samples that I made here, I ran UV Vis. I noted that the tube for the 150 sample was tinged a slight purple color, indicating that some AuNPs had adsorbed to its surface. I expect a slight diminishing of the signal for both UVVis and AA for this sample.


AA - uncentrifuged

Using the same samples as above ... I also made 6 standard concentrations for a calibration curve

  1. 0
    1. 10mL of water
  2. 10
    1. 100uL of 1000±10ug/mL Au standard into a 10mL volumetric flask with water
  3. 20
    1. 200uL of 1000±10ug/mL Au standard into a 10mL volumetric flask with water
  4. 30
    1. 300uL of 1000±10ug/mL Au standard into a 10mL volumetric flask with water
  5. 40
    1. 400uL of 1000±10ug/mL Au standard into a 10mL volumetric flask with water
  6. 50
    1. 500uL of 1000±10ug/mL Au standard into a 10mL volumetric flask with water

Here is the data for these samples

UVVis - centrifuged

I centrifuged 1mL of each sample in the centrifuge in 207 for 30 minutes at 13200rpm.