User:Melissa Novy/Notebook/CHEM-572/2013/02/13

From OpenWetWare

< User:Melissa Novy | Notebook | CHEM-572 | 2013 | 02(Difference between revisions)
Jump to: navigation, search
(Autocreate 2013/02/13 Entry for User:Melissa_Novy/Notebook/CHEM-572)
Current revision (11:02, 27 February 2013) (view source)
(Objectives)
 
(5 intermediate revisions not shown.)
Line 15: Line 15:
<!-- ##### DO NOT edit above this line unless you know what you are doing. ##### -->
<!-- ##### DO NOT edit above this line unless you know what you are doing. ##### -->
-
==Entry title==
+
==Objectives==
-
* Insert content here...
+
* Make a calibration curve for the determination of Ag<sup>+</sup> in solutions.
 +
** Make 0.10 M AgNO<sub>3</sub> solution.
 +
* Remove PLA2002D + 5 wt % LMT film from glass mold.
 +
* Prepare a starter culture of DH5α-T1 cells.
 +
** Please refer to [[User:Keyun_Wang/Notebook/Experimental_Biological_Chemistry_I/2013/02/13|Keyun Wang's entry]] for the protocol.
 +
==0.1 M AgNO<sub>3</sub> Solution==
 +
* Calculations
 +
  '''250 mL of 0.1 M AgNO<sub>3</sub>'''
 +
  '''MW AgNO<sub>3</sub> = 169.87 g/mol'''
 +
  '''0.1 M × 0.250 L = 0.025 mol'''
 +
  '''0.025 mol AgNO<sub>3</sub> × (169.87 g AgNO<sub>3</sub>/1 mol AgNO<sub>3</sub>) = 4.24675 g AgNO<sub>3</sub>'''
-
<!-- ##### DO NOT edit below this line unless you know what you are doing. ##### -->
+
* Actual mass AgNO<sub>3</sub>: 4.24694 g
 +
* Actual molarity AgNO<sub>3</sub>: 0.10000 M AgNO<sub>3</sub>
 +
 
 +
==Ag<sup>+</sup> Ion Selective Electrode Calibration==
 +
* The protocol from [[User:Melissa_Novy/Notebook/CHEM-572/2013/02/12|2013/02/12]] was followed using solutions of the following [AgNO<sub>3</sub>]:
 +
** 0 M AgNO<sub>3</sub>
 +
** 0.0001 M AgNO<sub>3</sub>
 +
** 0.0010 M AgNO<sub>3</sub>
 +
** 0.0105 M AgNO<sub>3</sub>
 +
** 0.1000M AgNO<sub>3</sub>
 +
 
 +
* Conductivity of solutions of AgNO<sub>3</sub>
 +
 
 +
{| {{table}}
 +
| align="center" style="background:#f0f0f0;"|'''[Ag<sup>+</sup>] [M]'''
 +
| align="center" style="background:#f0f0f0;"|'''log [Ag<sup>+</sup>] []'''
 +
| align="center" style="background:#f0f0f0;"|'''Conductivity [mV]'''
 +
|-
 +
| 0.00000||0||394.6
 +
|-
 +
| 0.00000||0||390.7
 +
|-
 +
| 0.00000||0||386.4
 +
|-
 +
| 0.00010||-4||377
 +
|-
 +
| 0.00010||-4||375.2
 +
|-
 +
| 0.00010||-4||374.7
 +
|-
 +
| 0.00100||-3||411.6
 +
|-
 +
| 0.00100||-3||411.8
 +
|-
 +
| 0.00100||-3||411.8
 +
|-
 +
| 0.01050||-1.98||470.1
 +
|-
 +
| 0.01050||-1.98||470.3
 +
|-
 +
| 0.01050||-1.98||470.3
 +
|-
 +
| 0.10000||-1||528.5
 +
|-
 +
| 0.10000||-1||528.5
 +
|-
 +
| 0.10000||-1||528.5
|}
|}
 +
 +
[[Image:AgCC.png]]
 +
 +
* The error bars indicate a 5% deviation.
 +
<br>
 +
 +
* Conductivity of filtrate solutions
 +
{| {{table}}
 +
| align="center" style="background:#f0f0f0;"|'''Sample []'''
 +
| align="center" style="background:#f0f0f0;"|'''Conductivity [mV]'''
 +
| align="center" style="background:#f0f0f0;"|'''Average [mV]'''
 +
| align="center" style="background:#f0f0f0;"|'''log [Ag+] []'''
 +
| align="center" style="background:#f0f0f0;"|'''[Ag+] [M]'''
 +
|-
 +
| (50:50) H2O/EtOH||372||374.3333333||-3.894305004||0.000127554
 +
|-
 +
| ||377.8||||||
 +
|-
 +
| ||373.2||||||
 +
|-
 +
| [[User:Melissa_Novy/Notebook/CHEM-572/2013/02/12|100Ag-LMT Filtrate]]||403.6||389.7333333||-3.595826469||0.000253614
 +
|-
 +
| ||389.3||||||
 +
|-
 +
| ||376.3||||||
 +
|-
 +
| [[User:Melissa_Novy/Notebook/CHEM-572/2013/01/30|95Ag-LMT Filtrate]]||582.3||582.7666667||0.145492134||1.397951596
 +
|-
 +
| ||582.9||||||
 +
|-
 +
| ||583.1||||||
 +
|}
 +
 +
<!-- ##### DO NOT edit below this line unless you know what you are doing. ##### -->
 +
__NOTOC__
__NOTOC__

Current revision

Search this Project

Customize your entry pages

Objectives

  • Make a calibration curve for the determination of Ag+ in solutions.
    • Make 0.10 M AgNO3 solution.
  • Remove PLA2002D + 5 wt % LMT film from glass mold.
  • Prepare a starter culture of DH5α-T1 cells.

0.1 M AgNO3 Solution

  • Calculations
  250 mL of 0.1 M AgNO3
  MW AgNO3 = 169.87 g/mol
  0.1 M × 0.250 L = 0.025 mol
  0.025 mol AgNO3 × (169.87 g AgNO3/1 mol AgNO3) = 4.24675 g AgNO3
  • Actual mass AgNO3: 4.24694 g
  • Actual molarity AgNO3: 0.10000 M AgNO3

Ag+ Ion Selective Electrode Calibration

  • The protocol from 2013/02/12 was followed using solutions of the following [AgNO3]:
    • 0 M AgNO3
    • 0.0001 M AgNO3
    • 0.0010 M AgNO3
    • 0.0105 M AgNO3
    • 0.1000M AgNO3
  • Conductivity of solutions of AgNO3
[Ag+] [M] log [Ag+] [] Conductivity [mV]
0.000000394.6
0.000000390.7
0.000000386.4
0.00010-4377
0.00010-4375.2
0.00010-4374.7
0.00100-3411.6
0.00100-3411.8
0.00100-3411.8
0.01050-1.98470.1
0.01050-1.98470.3
0.01050-1.98470.3
0.10000-1528.5
0.10000-1528.5
0.10000-1528.5

Image:AgCC.png

  • The error bars indicate a 5% deviation.


  • Conductivity of filtrate solutions
Sample [] Conductivity [mV] Average [mV] log [Ag+] [] [Ag+] [M]
(50:50) H2O/EtOH372374.3333333-3.8943050040.000127554
377.8
373.2
100Ag-LMT Filtrate403.6389.7333333-3.5958264690.000253614
389.3
376.3
95Ag-LMT Filtrate582.3582.76666670.1454921341.397951596
582.9
583.1



Personal tools