Objectives
- Perform error analysis on the compiled data calibration curve.
- Determine the standard deviation for the data points.
- Determine the confidence interval for 90% and 95% confidence.
- Determine if any data can be ruled out using a Q-test.
- Determine the molar extinction coefficient for both adenosine and inosine.
- Determine the concentration of an unknown adenosine sample(taken from Javier's group).
- Perform Grubb's test on outlying peak of adenosine molar extinction coefficent graph.
- Perform data analysis on molar extinction coefficient graphs(both adenosine and inosine).
Group calibration curve data
Figure 1. CHEM-571 Calibration Curve for Adenosine
- The Grubb's test will be employed to determine whether the points that lie outside of the line of best fit are outliers
Figure 2. CHEM-571 Calibration Curve for Inosine
Table 1. Molar absorptivity for Adenosine and Inosine
Molar Absorptivity(Lmol-1cm-1)
|
'
|
Adenosine |
14025
|
Inosine |
11007
|
Table 2. Standard deviation of Inosine Absorbance values
Comcentration(M)
|
Standard Deviation
|
4.80E-05 |
0.10707708
|
4.00E-05 |
0.12254917
|
3.20E-05 |
0.0943149
|
2.40E-05 |
0.06671357
|
1.60E-05 |
0.04841797
|
8.00E-06 |
0.04748368
|
Table 3. Standard deviation of Adenosine Absorbance values
Concentration(M)
|
Standard Deviation
|
3.00E-05 |
0.072537576
|
2.50E-05 |
0.045680412
|
2.00E-05 |
0.064123319
|
1.50E-05 |
0.056753854
|
1.00E-05 |
0.0313289
|
5.00E-06 |
0.046559639
|
Table 4. 95% Confidence Intervals for Adenosine Absorbance values
Concentration(M)
|
Confidence Interval
|
3.00E-05 |
0.063580821
|
2.50E-05 |
0.04003991
|
2.00E-05 |
0.056205534
|
1.50E-05 |
0.049746033
|
1.00E-05 |
0.027460487
|
5.00E-06 |
0.040810573
|
Table 5. 90% Confidence Intervals for Adenosine Absorbance Values
Concentration(M)
|
Confidence Interval
|
3.00E-05 |
0.053358707
|
2.50E-05 |
0.033602552
|
2.00E-05 |
0.047169171
|
1.50E-05 |
0.041748187
|
1.00E-05 |
0.023045567
|
5.00E-06 |
0.034249313
|
Table 6. 95% Confidence Intervals for Inosine Absorbance Values
Concentration(M)
|
Confidence Interval
|
4.80E-05 |
0.093855474
|
4.00E-05 |
0.10741711
|
3.20E-05 |
0.082669136
|
2.40E-05 |
0.058475948
|
1.60E-05 |
0.042439442
|
8.00E-06 |
0.041620516
|
Table 7. 90% Confidence Intervals for Inosine Absorbance Values
Concentration(M)
|
Confidence Interval
|
4.80E-05 |
0.078765997
|
4.00E-05 |
0.09014728
|
3.20E-05 |
0.069378126
|
2.40E-05 |
0.049074562
|
1.60E-05 |
0.035616303
|
8.00E-06 |
0.034929038
|
Unknown Adenosine Solution
Figure 3. Absorption Spectrum of Unknown Concentration
- Using the molar extinction coefficient determined from the calibration curve above the concentration of the unknown is 4.7344E-5 M.
|