Line 50: | Line 50: | ||
# Remove the electrode from its storage solution. | # Remove the electrode from its storage solution. | ||
# Rinse the end of the electrode with '''distilled''' H<sub>2</sub>O | # Rinse the end of the electrode with '''distilled''' H<sub>2</sub>O | ||
# Dry the end of the electrode with a | # Dry the end of the electrode with a kimwipe(do not use a paper towel!) | ||
# Make sure not to set the electrode down | # Make sure not to set the electrode down | ||
# Place the electrode directly into the solution to be analyzed. | # Place the electrode directly into the solution to be analyzed. | ||
Line 57: | Line 57: | ||
# Once the measurement is finalized, record the value that appears on the screen. | # Once the measurement is finalized, record the value that appears on the screen. | ||
<br> | <br> | ||
The general protocol for ion selective electrode was used for | The general protocol for ion selective electrode was used for the collection of the ion selective electrode measurements | ||
*Upon testing the pH of our standard solutions, it was observed that the pH was too low(~7.8). | |||
*To adjust the pH of the solution, 0.1mL of ionic strength adjuster (4M pH adjuster containing NaOH) was added to each of the 3mL solutions before testing. | |||
*The ionic strength pH adjuster adjusted the pH of the solutions to a new pH value between 8 and 10. | |||
Revision as of 20:13, 4 October 2012
Project name | <html><img src="/images/9/94/Report.png" border="0" /></html> Main project page <html><img src="/images/c/c3/Resultset_previous.png" border="0" /></html>Previous entry<html> </html>Next entry<html><img src="/images/5/5c/Resultset_next.png" border="0" /></html> | ||||||||||||||||||||||||
Objectives
Crosslinking
Ion Selective ElectrodeGeneral Protocol:
|