#Edit calibration curve to make sure it is in ppm and first order | #Edit calibration curve to make sure it is in ppm and first order | ||
#Make sure that the Fuel gas parameter is at 1.8L/min and that the Support gas is at 8L/min | #Make sure that the Fuel gas parameter is at 1.8L/min and that the Support gas is at 8L/min | ||
#Select Next. Then double click on calibration curve, reenter calibration and enter the number of standard solutions created | #Select Next. Then double click on calibration curve, reenter calibration and enter the number of standard solutions created(standards should be entered from lowest concentration to highest concentration). Click OK | ||
# | #Click Next | ||
#Label all samples | |||
#Select "Collect Send" | |||
#Instrument Checklist box will appear. Click all boxes | |||
#Hold the purge button on the machine and make sure that the air and fuel level are at the right level(1,8,8) | |||
#Click line search, close, next, and next | |||
#Click finish. | |||
#Hold the purge and ignite button on the AA until a flame appears | |||
#Allow the sample tube to collect deionized water for autozero. Click autozero | |||
#After autozero is complete, test labeled samples in the order in which they were entered | |||
**Note: Between each sample, allow the sample collecting tube to collect deionized H<sub>2</sub>O until absorption value on the computer screen in close to 0. Then able to collect sample | |||
==AA results from CuSO<sub>4</sub> solutions== | |||
Gold nano particles were clogging the sample collection tube. All results were invalid. The data will be recollected on the next date | |||
Revision as of 18:17, 14 October 2012
Project name | <html><img src="/images/9/94/Report.png" border="0" /></html> Main project page <html><img src="/images/c/c3/Resultset_previous.png" border="0" /></html>Previous entry<html> </html>Next entry<html><img src="/images/5/5c/Resultset_next.png" border="0" /></html> | |||||||||||||||||||||||||||||||||||
Objectives
*NaSO4 solutions were prepared 2012/9/21 CrosslinkingNote: Films crosslinked on 10/03/2012 still remained intact in deionized H2O despite using too little glutaraldehyde.
Note: All GA films were cut in half prior to crosslinking. Remaining uncrosslinked glutaraldehyde films remained under the fume hood
Atomic AbsorptionGeneral Protocol for use of AA-6200 Shimadzu:
AA results from CuSO4 solutionsGold nano particles were clogging the sample collection tube. All results were invalid. The data will be recollected on the next date
|