Bioruptor

From OpenWetWare

Jump to: navigation, search

Bioruptor (B435)

BEFORE USE:

1. New user MUST be trained by the captain or present users.


INSTRUCTION FOR USE:

DO NOT turn on the instrument when the chiller is not filled with water

DO NOT exceed 1 hr total sonication time. Allow 20 minutes of cooling between samples

Optimal bath temperature is 4°C. Sample temperature should not exceed 10°C

Nominal sample volumes:

[1] 0.6 mL ependorf tube = 50 - 100 μL

[2] 1.5 mL ependorf tube = 100 - 300 μL

[3] 15 mL conical tube = 0.5 - 2 mL


PLEASE REPORT PROBLEMS TO GIOVANNI LUCHETTI OR RAMIN DUBEY IN B435

WEBSITE: http://www.diagenode.com/en/catalog/shearing-technologies-54/bioruptor--55/product/bioruptor-plus-sonication-device-with-1-5-15-ml-tube-holder-2

MANUFACTURER: DIAGENODE