User:Paul Rothenberg/Notebook/CHEM 571 Fall 2014/2014/10/22

From OpenWetWare

Jump to: navigation, search
Project name Main project page
Previous entry      Next entry

KI Electrode Testing

We continued with making an I- ISE to determine concentration of iodide in solution.

Procedure

  1. Ag/AgCl Electrode:
    1. The plastic covering was removed, and stored in the 3 M NaCl solution
  2. Double-Junction:
    1. Diluted saturated KNO3 was added to the double junction cell
  3. Silver Wire:
    1. A piece was cut and placed in KI/KNO3 solution and electrolyzed
  4. KI Standard Curve:
    1. The Ag/AgCl electrode was placed in the double-junction electrode. The Ag/AgCl electrode was attached to the multimeter
    2. The silver wire was attached directly to the alligator clip
    3. KI solutions were tested for potential


Results


[KI], mM ln of KI Potential, mV
503.912023005-278.3
51.609437912-223
0.5-0.693147181-161.8
0.05-2.995732274-105.1
0.005-5.298317367-70