Revision as of 17:15, 2 June 2010

Flow Rates

Table 2-3. Effective flow rate range SilicaTip emitters (from Finnigan Nanospray Ion Source Operator’s Manual 97055-97025 Revision A, October 2004)

Emitter ID (um) Flow rate (nL/min)
5 20 to 100
8 100 to 300
10 100 to 400
15 150 to 400
30 300 to 1000


Table 9-1. LC/NSI/MS Parameters for Finnigan LTQ or LCQ Series mass spectrometers (Finnigan™ Nanospray Ion Source Operator’s Manual 97055-97025 Revision A October 2004)

Column Size Flow Rate Tip ID Emitter Position Capillary Temperature
50 μm <100 to 300 nL/min 5 to 15 μm approx. 1 mm 100 to 180 °C
75 μm 150 to 800 nL/min 10 to 20 μm 1 to 2 mm 130 to 200 °C
180 μm 0.3 to 1.0 μL/min 20 to 30 μm 2 to 3 mm 150 to 250 °C
320 μm 1 to 5 μL/min 30 to 75 μm 3 to 5 mm 150 to 250 °C

This is from the Thermo Chromatography Catalogue 2006-2007 (pg. 177).

ColumnID (mm) Flow Rate Range (uL/min) Optimum Flow Rate (ul/min) Recommended Injection Vol. (ul) API Source
0.32 4 - 10 5 0.15 Micro-ESI
0.18 1 - 3 2 0.05 Micro-ESI
0.1 0.4 - 1 0.5 0.015 Nanospray
0.075 0.2 - 0.5 0.3 0.01 Nanospray