Data was gathered from both the Spree technology and the Gold Standards for measuring both Heart rate and temperature. The mean values from the Spree technology used to measure heart rate differed by approximately 6 beats per minute from the gold standard pulse ox measurement. The mean values from the Spree technology used to measure body temperature differed by approximately 1 degree Fahrenheit. These values however do not properly represent the accuracy of the Spree device in measuring heart rate and temperature. By analyzing the correlation coefficients and the values obtained from a t-test, heart rate was found to be accurately measured by the Spree technology, while body temperature values were significantly different than the values obtained from the gold standard. This data supports the conclusion that the Spree technology may be used to accurately measure heart rate but not body temperature.





Revision as of 10:01, 20 September 2017

BME 100 Fall 2017 Home
People
Lab Write-Up 1 | Lab Write-Up 2 | Lab Write-Up 3
Lab Write-Up 4 | Lab Write-Up 5 | Lab Write-Up 6
Course Logistics For Instructors
Photos
Wiki Editing Help


OUR TEAM

Name: Caiden Gledhill
Name: Erin Lockwood
Name: Alexandra Richey
Name: Maryah Alalwan
Name: Matthew Tam
Name: Aymard Irakoze

LAB 3 WRITE-UP

Descriptive Stats and Graph

Heart Rate Means:

Gold Standard: 98.08977 beats/minute

Spree Data: 92.0 beats/minute

Heart Rate Standard Deviation:

Gold Standard: 11.31371

Spree Data: 14.14214

Temperature Means:

Gold Standard: 96.64716 degrees Fahrenheit

Spree Data: 95.53086 degrees Fahrenheit

Temperature Standard Deviation:

Gold Standard: 1.9226

Spree Data: 0.870378




Inferential Stats

Heart Rate Correlation Coefficient: 0.690806

Temperature Correlation Coefficient: 0.192798089

T-Test Heart Rate: 0.427116

T-Test Temperature: 1.09676e^-21

Graphing

Heart Rate:

Temperature:

Summary of Results



Data was gathered from both the Spree technology and the Gold Standards for measuring both Heart rate and temperature. The mean values from the Spree technology used to measure heart rate differed by approximately 6 beats per minute from the gold standard pulse ox measurement. The mean values from the Spree technology used to measure body temperature differed by approximately 1 degree Fahrenheit. These values however do not properly represent the accuracy of the Spree device in measuring heart rate and temperature. By analyzing the correlation coefficients and the values obtained from a t-test, heart rate was found to be accurately measured by the Spree technology, while body temperature values were significantly different than the values obtained from the gold standard. This data supports the conclusion that the Spree technology may be used to accurately measure heart rate but not body temperature.




Experimental Design of Own Device

Group Size: (Maximize if possible)

Sample Size: Should be Americans, (Near Maryland since it's near the FDA Office???) since our goal is to bring our device to the United States through the FDA

Need a blood sampling to compare to our product (Gold standard)

Needs to be over time to compare differences?

Should take multiple samples of our test to test the reliability

Procedure: