Knight:Dialysis: Difference between revisions

From OpenWetWare
Jump to navigationJump to search
No edit summary
 
Line 12: Line 12:
#'''Optional:''' to remove contaminating glycerol from the dialysis unit, dialyze the unit in 1L of DI water for 15 mins.  (Glycerol content: <3% in 3K, ~15% in 7K and ~23% in 10K MINI)
#'''Optional:''' to remove contaminating glycerol from the dialysis unit, dialyze the unit in 1L of DI water for 15 mins.  (Glycerol content: <3% in 3K, ~15% in 7K and ~23% in 10K MINI)
#'''Optional:''' to remove contaminating metals, dialyze 15 minutes against 1 L 1 mM EDTA.  (Metals present in a 3K, 7K or 10K MINI; 2 ppb iron, 5 ppb magnesium, 1.5 ppb nickel, 0.2 ppb zinc, 0.2 ppb copper, 0.5 ppb chromium and 0.3 ppb cadmium)
#'''Optional:''' to remove contaminating metals, dialyze 15 minutes against 1 L 1 mM EDTA.  (Metals present in a 3K, 7K or 10K MINI; 2 ppb iron, 5 ppb magnesium, 1.5 ppb nickel, 0.2 ppb zinc, 0.2 ppb copper, 0.5 ppb chromium and 0.3 ppb cadmium)
#*The volume level of the sample should be higher than the dialysate to avoid hydrostatic pressure forcing dialysate into the unit (thereby diluting the sample).
#Apply sample with a standard pipette.
#Apply sample with a standard pipette.
#*Sample volume should be between 10-100 &mu;L.
#*''Sample volume should be between 10-100 &mu;L.''
#Cap the dialysis unit and place in a floatation device.
#Cap the dialysis unit and place in a floatation device.
#Add 0.5-1L of dialysate to beaker.
#Add 0.5-1L of dialysate to beaker.
Line 20: Line 19:
#Place the beaker in ice or in a cold room and on a stir plate.
#Place the beaker in ice or in a cold room and on a stir plate.
#Place the float with dialysis unit in the beaker so that the bottom of the dialysis unit is in contact with the dialysate.
#Place the float with dialysis unit in the beaker so that the bottom of the dialysis unit is in contact with the dialysate.
#*''The volume level of the sample should be higher than the dialysate to avoid hydrostatic pressure forcing dialysate into the unit (thereby diluting the sample).''
#Use a low speed setting on the stir plate to avoid submerging the unit.
#Use a low speed setting on the stir plate to avoid submerging the unit.
#Equilibrate for 2 hours.
#Equilibrate for 2 hours.
#Collect the sample from the corner of the Slide-A-Lyzer unit.
#Collect the sample from the corner of the Slide-A-Lyzer unit.

Revision as of 15:38, 19 September 2006

in progress!

Purpose

To change buffers in which your protein resides.

Materials

Procedure

  1. Wear gloves to prevent contamination.
  2. Optional: to remove contaminating glycerol from the dialysis unit, dialyze the unit in 1L of DI water for 15 mins. (Glycerol content: <3% in 3K, ~15% in 7K and ~23% in 10K MINI)
  3. Optional: to remove contaminating metals, dialyze 15 minutes against 1 L 1 mM EDTA. (Metals present in a 3K, 7K or 10K MINI; 2 ppb iron, 5 ppb magnesium, 1.5 ppb nickel, 0.2 ppb zinc, 0.2 ppb copper, 0.5 ppb chromium and 0.3 ppb cadmium)
  4. Apply sample with a standard pipette.
    • Sample volume should be between 10-100 μL.
  5. Cap the dialysis unit and place in a floatation device.
  6. Add 0.5-1L of dialysate to beaker.
  7. Add stir bar.
  8. Place the beaker in ice or in a cold room and on a stir plate.
  9. Place the float with dialysis unit in the beaker so that the bottom of the dialysis unit is in contact with the dialysate.
    • The volume level of the sample should be higher than the dialysate to avoid hydrostatic pressure forcing dialysate into the unit (thereby diluting the sample).
  10. Use a low speed setting on the stir plate to avoid submerging the unit.
  11. Equilibrate for 2 hours.
  12. Collect the sample from the corner of the Slide-A-Lyzer unit.