==UV-Vis standards==
==UV-Vis standards==
* Insert content here...
* Standards needed to be made of R6G in order to calculate the amount of dye that was absorbed and how much dye remains in the solution.  
 
**Molar Absorptivity of R6G: 116000 LMol<sup>-1</sup>cm<sup>-1</sup>
Assumed stock concentration: .005 M
 
**Because the absorbance of R6G is so high, standard concentrations needed to be diluted to below 8 μM.
 
Table 1. Standard concentrations in 2mL solutions
 
{| {{table}}
| align="center" style="background:#f0f0f0;"|'''Standard Concentrations (uM)'''
| align="center" style="background:#f0f0f0;"|'''Actual Concentration (M)'''
| align="center" style="background:#f0f0f0;"|'''Absorbance'''
|-
| 0||0||0
|-
| 2||1.59483E-06||0.185
|-
| 4||3.18966E-06||0.37
|-
| 6||4.78448E-06||0.555
|-
| 8||6.37931E-06||0.74
|}
 
"The concentration of the original stock solution was 398.7μM
 
Table 2: Actual concentration of dye in eppendorf samples
 
{| {{table}}
| align="center" style="background:#f0f0f0;"|'''Sample number'''
| align="center" style="background:#f0f0f0;"|'''believed concentration'''
| align="center" style="background:#f0f0f0;"|'''Concentration in 1mL solution (uM)'''
| align="center" style="background:#f0f0f0;"|'''molar concentration in epp tubes (M)'''
|-
| 1||20||15.94827586||1.59483E-05
|-
| 2||40||31.89655172||3.18966E-05
|-
| 3||60||47.84482759||4.78448E-05
|-
| 4||80||63.79310345||6.37931E-05
|-
| 5||100||79.74137931||7.97414E-05
|-
| 6||150||119.612069||0.000119612
|-
| 7||200||159.4827586||0.000159483
|}
 
 
 
 
*Notes*
-These samples were left for too long without being tested and resulted in some of the dye releasing out of the complexes. For this reason, UV-Vis of these samples were not fully run and new samples were made.  





Revision as of 15:39, 23 September 2014

Project name <html><img src="/images/9/94/Report.png" border="0" /></html> Main project page
<html><img src="/images/c/c3/Resultset_previous.png" border="0" /></html>Previous entry<html>&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;</html>Next entry<html><img src="/images/5/5c/Resultset_next.png" border="0" /></html>

UV-Vis standards

  • Standards needed to be made of R6G in order to calculate the amount of dye that was absorbed and how much dye remains in the solution.
    • Molar Absorptivity of R6G: 116000 LMol-1cm-1

Assumed stock concentration: .005 M

    • Because the absorbance of R6G is so high, standard concentrations needed to be diluted to below 8 μM.

Table 1. Standard concentrations in 2mL solutions

Standard Concentrations (uM) Actual Concentration (M) Absorbance
0 0 0
2 1.59483E-06 0.185
4 3.18966E-06 0.37
6 4.78448E-06 0.555
8 6.37931E-06 0.74

"The concentration of the original stock solution was 398.7μM

Table 2: Actual concentration of dye in eppendorf samples

Sample number believed concentration Concentration in 1mL solution (uM) molar concentration in epp tubes (M)
1 20 15.94827586 1.59483E-05
2 40 31.89655172 3.18966E-05
3 60 47.84482759 4.78448E-05
4 80 63.79310345 6.37931E-05
5 100 79.74137931 7.97414E-05
6 150 119.612069 0.000119612
7 200 159.4827586 0.000159483



  • Notes*

-These samples were left for too long without being tested and resulted in some of the dye releasing out of the complexes. For this reason, UV-Vis of these samples were not fully run and new samples were made.